Fluke 7350 specifications Menu Program Edit Menu

Page 43

Local Operation 4

Menu

3.Use the rotary knob to highlight “new”.

4.Press Edit [F4]. Since “new” was highlighted, the Calibrator will create a new program and give a default name of NAMEnn, where nn is a two digit number. The Program editing screen will appear, displaying the first step.

gld19.bmp

Figure 4-17. Menu Program Edit - Menu

5.Using the numeric keypad, enter the values for Pressure, Tolerance, Dwell time, and Max time, pressing ENTER after each value. The rotary knob may be used to skip fields.

6.At the 0 step, enter the next pressure and the Calibrator will automatically insert a new pressure step.

7.Repeat steps 5 and 6 until the test sequence is complete. The Tolerance, Dwell time and Max time will default to the value programmed in the first program step, and only need to be changed if a different value is desired for a particular step.

8.It is often suggested that the last point in a calibration program is to vent the pressure to atmosphere. This can be achieved by selecting the VENT function key as the pressure set point.

Note

You can use the vent as a set-point at any point in the program. The dwell and max time function with vent just as they would at a controlled set-point.

9.When all steps have been entered, press PREVIOUS to return to the Named programs screen. See section, Chapter 4, Changing the Name of a Program, section for creating a user-defined program name.

4-19

Image 43
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii General Information How to Contact FlukeGeneral Information IntroductionSafety Information Features Symbols Used in this ManualEasily Programmable Power-On Self TestAutomatic Zero Adjust Ease of OperationRuska 7350 Options List Part Number Standard Equipment & OptionsRuska Back Plane Board Power SupplyTheory of Operation Electronics SectionRuska 7350 Block Diagram Front Panel IEEE-488 InterfaceMotor Controller Gearmotor Pneumatic Module76xx Interface Board Pneumatic PumpTest Port Electronics RegulatorDome Regulator System TransducerRuska Unpacking the Ruska InstallationPneumatic Connections Powering Up the RuskaDrive Air Supply Port Function Keys Local OperationLocal Operation Numeric KeypadMenu Tree Cancel, PreviousMain Menu TutorialMain Menu Entering Pressure Tare Mode Selecting Mode of OperationSelecting Language Triple Range Optional Setting the Pressure Set-PointMain Menu Controlling PressureStep/Jog VentMenu Menu SetupAuto Vent Menu Setup LimitsHigh Limit Low LimitControl Band Slew Limit AccessMenu Setup User Slew RateGas Head Pressure Correction Step SizeBar Graph Maximum Ready ToleranceKey Click Menu Setup UnitsPressure Filter Changing the Number of Decimals13. Menu Setup Units Edit Name Menu Protocol Menu Setup RemoteMenu Setup System Gpib AddressMenu Calibrate ResetCalibration Password Date/TimePreparing to Program Menu Program Storing a Sequence in MemoryMaxTime Mode of EntryEntering a New Program 17. Menu Program Edit Menu 18. Menu Program Edit Auto Menu Automatically Generating a ProgramChanging an Existing Program Changing the Name of a Program20. Menu Program Edit Menu Changing the Configuration Stored with a ProgramRunning a Program 21. Menu Program Config MenuMenu Test Sweep Test Menu TestMenu Test Remote Test Menu Test Self Test26. Menu Test Shop1 Menu Menu Test Shop1Menu Display Blank Menu DisplayRuska IEEE-488 Remote OperationRemote Operation CapabilitiesRS-232 Remote/Local OperationScpi Response Format ConfigurationScpi Command Format Device MessagesScpi Command Summary ANSI/IEEE 488.2-1987 Command SummaryMODE? List Example Scpi CommandsScpi Status Registers Interface Panel Emulation Serial Operation Ruska Observing the Software Version Number MaintenancePreventive Maintenance Observing the Ruska 7350’s Full Scale RatingCalibration Processor BatteryPreparation Calibration Instructions Single SensorStoring the Coefficients Zeroing Editing the Calibration CoefficientsRuska 7350 Pump Lubrication CleaningFuse Disconnection Instructions Preparation for Storage and ShipmentPacking Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Packing the Calibrator Shipping Instructions7350 Accuracy Summary of SpecificationsSpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range