Local Operation 4
Menu
3.Use the numeric keypad to enter the new calibration password. Setting the calibration password to zero allows free access to Calibrator calibration and constants. Press
ENTER.
4.Press Yes [F4] to acknowledge changing the calibration password. Press No [F5] to reject changing the calibration password.
Menu Program — Storing A Sequence in Memory
Consider an exercise that requires the Calibrator to start at 30 psi, increase to 50 psi, then decrease to 20 psi. Test sequences like this may be stored in the Calibrator’s memory as a program.
One benefit of storing a sequence in memory is that the user does not have to command each pressure separately every time a multipoint test or calibration routine is performed. Another benefit is that the user can specify a tolerance for each
Note
The Calibrator can store up to 1,000 program steps that may be divided among a maximum of 20 named programs.
Preparing to Program
Before entering a test sequence, consider the items discussed below.
Program Name
Valid program names range from one to eight characters in length and can include numbers, upper case letters, and the /, %, and # symbols. For example, Exer#14 and %FS test are both valid names.
Configuration
Since a program will depend on the current setup of the Calibrator (the current configuration is stored with the program) the user should set the units, limits, control parameters, etc., to the desired values before creating a program.
Number of Set-points
Before entering the sequence, the user should determine the number of upscale and downscale
Set-point Pressure and Tolerance
Each
(0.35 kPa), whereas another
Dwell Time
Once the pressure is within the specified tolerance, the Calibrator starts a timer that runs for a
Usually dwell time has a value of a few seconds, but a value of 0 can be used to create a pause in the program. When the dwell time is set to zero, the Calibrator switches to manual control once it is within the tolerance value of the