Fluke 7350 specifications Menu Program Storing a Sequence in Memory, Preparing to Program

Page 41

Local Operation 4

Menu

3.Use the numeric keypad to enter the new calibration password. Setting the calibration password to zero allows free access to Calibrator calibration and constants. Press

ENTER.

4.Press Yes [F4] to acknowledge changing the calibration password. Press No [F5] to reject changing the calibration password.

Menu Program — Storing A Sequence in Memory

Consider an exercise that requires the Calibrator to start at 30 psi, increase to 50 psi, then decrease to 20 psi. Test sequences like this may be stored in the Calibrator’s memory as a program.

One benefit of storing a sequence in memory is that the user does not have to command each pressure separately every time a multipoint test or calibration routine is performed. Another benefit is that the user can specify a tolerance for each set-point pressure. Once a tolerance is set, the dwell timer will start counting down instead of waiting until the pressure is at the exact set-point to start the dwell timer. This gives the user a time advantage compared to manual mode operation and provides a degree of automation without the use of an external computer.

Note

The Calibrator can store up to 1,000 program steps that may be divided among a maximum of 20 named programs.

Preparing to Program

Before entering a test sequence, consider the items discussed below.

Program Name

Valid program names range from one to eight characters in length and can include numbers, upper case letters, and the /, %, and # symbols. For example, Exer#14 and %FS test are both valid names.

Configuration

Since a program will depend on the current setup of the Calibrator (the current configuration is stored with the program) the user should set the units, limits, control parameters, etc., to the desired values before creating a program.

Number of Set-points

Before entering the sequence, the user should determine the number of upscale and downscale set-points required to complete the exercise.

Set-point Pressure and Tolerance

Each set-point in the program requires both a pressure and a tolerance, in the current units of measure. For example, one set-point might require a tolerance as low as 0.05 psi

(0.35 kPa), whereas another set-point in the same program could be satisfied with a tolerance as high as 5 psi (0.35 kPa).

Dwell Time

Once the pressure is within the specified tolerance, the Calibrator starts a timer that runs for a user-defined number of seconds. This is referred to as dwell time. As long as this timer is running, the Calibrator will remain at the designated set-point unless the max time (see below) elapses. When the dwell time expires, the Calibrator will proceed to the next step. Typically the dwell time should be set to a value less than the max time.

Usually dwell time has a value of a few seconds, but a value of 0 can be used to create a pause in the program. When the dwell time is set to zero, the Calibrator switches to manual control once it is within the tolerance value of the set-point pressure. The user must then press a key on the front panel to continue the program.

4-17

Image 41
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii General Information How to Contact FlukeIntroduction General InformationSafety Information Features Symbols Used in this ManualAutomatic Zero Adjust Power-On Self TestEase of Operation Easily ProgrammableRuska 7350 Options List Part Number Standard Equipment & OptionsRuska Theory of Operation Power SupplyElectronics Section Back Plane BoardRuska 7350 Block Diagram Motor Controller IEEE-488 InterfaceFront Panel 76xx Interface Board Pneumatic ModulePneumatic Pump GearmotorDome Regulator Electronics RegulatorSystem Transducer Test PortRuska Unpacking the Ruska InstallationDrive Air Supply Port Powering Up the RuskaPneumatic Connections Local Operation Local OperationNumeric Keypad Function KeysMenu Tree Cancel, PreviousMain Menu TutorialMain Menu Entering Pressure Selecting Language Selecting Mode of OperationTare Mode Main Menu Setting the Pressure Set-PointControlling Pressure Triple Range OptionalStep/Jog VentMenu Menu SetupHigh Limit Menu Setup LimitsLow Limit Auto VentControl Band Menu Setup User AccessSlew Rate Slew LimitBar Graph Maximum Step SizeReady Tolerance Gas Head Pressure CorrectionPressure Filter Menu Setup UnitsChanging the Number of Decimals Key Click13. Menu Setup Units Edit Name Menu Menu Setup System Menu Setup RemoteGpib Address ProtocolCalibration Password ResetDate/Time Menu CalibratePreparing to Program Menu Program Storing a Sequence in MemoryEntering a New Program Mode of EntryMaxTime 17. Menu Program Edit Menu 18. Menu Program Edit Auto Menu Automatically Generating a ProgramChanging an Existing Program Changing the Name of a Program20. Menu Program Edit Menu Changing the Configuration Stored with a ProgramRunning a Program 21. Menu Program Config MenuMenu Test Sweep Test Menu TestMenu Test Remote Test Menu Test Self Test26. Menu Test Shop1 Menu Menu Test Shop1Menu Display Blank Menu DisplayRuska Remote Operation Remote OperationCapabilities IEEE-488RS-232 Remote/Local OperationScpi Command Format ConfigurationDevice Messages Scpi Response FormatScpi Command Summary ANSI/IEEE 488.2-1987 Command SummaryMODE? List Example Scpi CommandsScpi Status Registers Interface Panel Emulation Serial Operation Ruska Preventive Maintenance MaintenanceObserving the Ruska 7350’s Full Scale Rating Observing the Software Version NumberCalibration Processor BatteryPreparation Calibration Instructions Single SensorStoring the Coefficients Zeroing Editing the Calibration CoefficientsFuse CleaningRuska 7350 Pump Lubrication Packing Instructions Preparation for Storage and ShipmentDisconnection Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Packing the Calibrator Shipping Instructions7350 Accuracy Summary of SpecificationsSpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range