Triple Range (Optional)

When Triple Ranging is enabled the currently selected range will be displayed below the current pressure. To change to a different range make sure the pressure is below 100 psi (700 kPa). Then press the Mode key and then press Range [F5] until the correct range is displayed. If the current pressure range is exceeded the range will automatically change to the highest range until manually changed.

Main Menu

The Main Menu displays the measured pressure in double-sized numbers. Below the pressure reading is the current unit of measure and mode (Tare or Absolute). The Main Menu can always be reached by repeatedly pressing PREVIOUS. The RUSKA 7350 is designed such that all of the commonly used functions are accessible by a direct key on the key pad or from a top level function key. Less commonly used set-up type functions are accessible through the menu function key.

gld09.bmp

Figure 4-7. Main Menu

Controlling Pressure

The Control [F2] function key is used to place the Calibrator into the Control Mode. A pressure set-point should be entered into the Calibrator prior to placing the unit into the control mode. This is to assure that the operator is aware of the set-point valve prior to entering into the control mode.

Setting the Pressure Set-Point

The pressure set-point is the destination of the pressure control algorithm. It should be set before entering Control mode. The pressure set-point is set to zero at power-up or whenever a pressure error occurs.

1.The pressure set-point is set from the Main Menu. (Press PREVIOUS until the Main Menu appears.)

2.Use the numeric keypad to enter the new pressure set-point in the current pressure units.

3.Press ENTER to accept the entry or press CLEAR to clear the numeric scratchpad.

4-6

Image 30
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii Introduction How to Contact FlukeGeneral Information General InformationSafety Information Symbols Used in this Manual FeaturesEase of Operation Power-On Self TestAutomatic Zero Adjust Easily ProgrammableStandard Equipment & Options Ruska 7350 Options List Part NumberRuska Electronics Section Power SupplyTheory of Operation Back Plane BoardRuska 7350 Block Diagram IEEE-488 Interface Front PanelMotor Controller Pneumatic Pump Pneumatic Module76xx Interface Board GearmotorSystem Transducer Electronics RegulatorDome Regulator Test PortRuska Installation Unpacking the RuskaPowering Up the Ruska Pneumatic ConnectionsDrive Air Supply Port Numeric Keypad Local OperationLocal Operation Function KeysCancel, Previous Menu TreeTutorial Main MenuMain Menu Entering Pressure Selecting Mode of Operation Tare ModeSelecting Language Controlling Pressure Setting the Pressure Set-PointMain Menu Triple Range OptionalVent Step/JogMenu Setup MenuLow Limit Menu Setup LimitsHigh Limit Auto VentControl Band Slew Rate AccessMenu Setup User Slew LimitReady Tolerance Step SizeBar Graph Maximum Gas Head Pressure CorrectionChanging the Number of Decimals Menu Setup UnitsPressure Filter Key Click13. Menu Setup Units Edit Name Menu Gpib Address Menu Setup RemoteMenu Setup System ProtocolDate/Time ResetCalibration Password Menu CalibrateMenu Program Storing a Sequence in Memory Preparing to ProgramMode of Entry MaxTimeEntering a New Program 17. Menu Program Edit Menu Automatically Generating a Program 18. Menu Program Edit Auto MenuChanging the Name of a Program Changing an Existing ProgramChanging the Configuration Stored with a Program 20. Menu Program Edit Menu21. Menu Program Config Menu Running a ProgramMenu Test Menu Test Sweep TestMenu Test Self Test Menu Test Remote TestMenu Test Shop1 26. Menu Test Shop1 MenuMenu Display Menu Display BlankRuska Capabilities Remote OperationRemote Operation IEEE-488Remote/Local Operation RS-232Device Messages ConfigurationScpi Command Format Scpi Response FormatANSI/IEEE 488.2-1987 Command Summary Scpi Command SummaryMODE? Example Scpi Commands ListScpi Status Registers Interface Panel Emulation Serial Operation Ruska Observing the Ruska 7350’s Full Scale Rating MaintenancePreventive Maintenance Observing the Software Version NumberProcessor Battery CalibrationCalibration Instructions Single Sensor PreparationStoring the Coefficients Editing the Calibration Coefficients ZeroingCleaning Ruska 7350 Pump LubricationFuse Preparation for Storage and Shipment Disconnection InstructionsPacking Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Shipping Instructions Packing the Calibrator7350 Summary of Specifications AccuracySpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range