Fluke 7350 specifications Storing the Coefficients

Page 66

RUSKA 7350

Users Manual

Note

If the actual pressure is outside of the tolerance for the requested mid-point pressure, Error –222 Data out of range will occur. Acknowledge this error by selecting OK, then re-enter the actual pressure, repeating Step 2, number 1, if necessary.

Step 3

1.To begin Step 3, use the calibration standard to Apply the high-point pressure requested by the 7350. The Measured pressure on the RUSKA 7350’s screen will change accordingly.

2.When the Measured pressure stabilizes, use the RUSKA 7350’s numeric keypad and OK to enter the actual pressure reported by the calibration standard. Do not enter the Measured pressure reported by the RUSKA 7350. If necessary, use the CLEAR key to correct a mistake in the edit field. If the actual pressure applied is acceptable, the Calibration, Step 4, screen will appear.

Note

If the actual pressure applied is outside of the tolerance for the requested high-point pressure, Error –222 Data out of range will occur. Acknowledge this error by selecting OK, then re-enter the actual pressure, repeating Step 3, number1, if necessary.

Step 4

1.To begin Step 4, use the calibration standard to Apply the mid-point pressure requested by the RUSKA 7350. As pressure is to be applied into the Test Port, the Measured pressure on the RUSKA 7350’s screen will change accordingly.

2.When the Measured pressure stabilizes, use the RUSKA 7350’s numeric keypad and OK to enter the actual pressure applied by the calibration standard. Do not enter the Measured pressure reported by the RUSKA 7350. If necessary, use the CLEAR key to correct a mistake in the edit field. If the actual pressure applied is acceptable, the Calibration complete screen will appear.

Storing the Coefficients

Note

In addition to saving the calibration coefficients to the RUSKA 7350’s memory, the user is advised to separately record the calibration coefficients and store this “backup” in a safe place.

Step 5

Calibration is complete. To exit the calibration procedure without storing the calibration coefficients in memory, press CANCEL. To store the calibration coefficients in memory, select DONE.

Step 6

Press PREV. to return to the main screen.

Once the calibration procedure is complete, the user is advised to record several pressure readings. If there are any significant errors at these points, then an error was probably made in generating one of the calibration pressures, and the calibration procedure should be repeated.

6-4

Image 66
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii Introduction How to Contact FlukeGeneral Information General InformationSafety Information Symbols Used in this Manual FeaturesEase of Operation Power-On Self TestAutomatic Zero Adjust Easily ProgrammableStandard Equipment & Options Ruska 7350 Options List Part NumberRuska Electronics Section Power SupplyTheory of Operation Back Plane BoardRuska 7350 Block Diagram IEEE-488 Interface Front PanelMotor Controller Pneumatic Pump Pneumatic Module76xx Interface Board GearmotorSystem Transducer Electronics RegulatorDome Regulator Test PortRuska Installation Unpacking the RuskaPowering Up the Ruska Pneumatic ConnectionsDrive Air Supply Port Numeric Keypad Local OperationLocal Operation Function KeysCancel, Previous Menu TreeTutorial Main MenuMain Menu Entering Pressure Selecting Mode of Operation Tare ModeSelecting Language Controlling Pressure Setting the Pressure Set-PointMain Menu Triple Range OptionalVent Step/JogMenu Setup MenuLow Limit Menu Setup LimitsHigh Limit Auto VentControl Band Slew Rate AccessMenu Setup User Slew LimitReady Tolerance Step SizeBar Graph Maximum Gas Head Pressure CorrectionChanging the Number of Decimals Menu Setup UnitsPressure Filter Key Click13. Menu Setup Units Edit Name Menu Gpib Address Menu Setup RemoteMenu Setup System ProtocolDate/Time ResetCalibration Password Menu CalibrateMenu Program Storing a Sequence in Memory Preparing to ProgramMode of Entry MaxTimeEntering a New Program 17. Menu Program Edit Menu Automatically Generating a Program 18. Menu Program Edit Auto MenuChanging the Name of a Program Changing an Existing ProgramChanging the Configuration Stored with a Program 20. Menu Program Edit Menu21. Menu Program Config Menu Running a ProgramMenu Test Menu Test Sweep TestMenu Test Self Test Menu Test Remote TestMenu Test Shop1 26. Menu Test Shop1 MenuMenu Display Menu Display BlankRuska Capabilities Remote OperationRemote Operation IEEE-488Remote/Local Operation RS-232Device Messages ConfigurationScpi Command Format Scpi Response FormatANSI/IEEE 488.2-1987 Command Summary Scpi Command SummaryMODE? Example Scpi Commands ListScpi Status Registers Interface Panel Emulation Serial Operation Ruska Observing the Ruska 7350’s Full Scale Rating MaintenancePreventive Maintenance Observing the Software Version NumberProcessor Battery CalibrationCalibration Instructions Single Sensor PreparationStoring the Coefficients Editing the Calibration Coefficients ZeroingCleaning Ruska 7350 Pump LubricationFuse Preparation for Storage and Shipment Disconnection InstructionsPacking Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Shipping Instructions Packing the Calibrator7350 Summary of Specifications AccuracySpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range