Automatically Generating a Program

For the Calibrator to automatically generate a program, the user must input the first set-point pressure, the last set-point pressure, and the number of steps in between, as well as the dwell time, max time and tolerance common to all set-points.

1.Ensure that the units, limits, and control parameters are set to their desired values.

2.The program is entered from the Program Menu. From the Main Menu (press PREVIOUS until the Main Menu appears), press Menu [F6] Program [F4].

3.Use the rotary knob to highlight “new”. If these steps are used on an existing program, all program steps will be deleted and replaced with the automatically generated program.

4.Press Edit [F4]. Since “new” was highlighted, the Calibrator will create a new program and give a default name of NAMEnn where nn is a two digit number. The Program editing screen will appear, displaying the first step.

5.Press Auto [F3].

gld20.bmp

Figure 4-18. Menu Program Edit Auto - Menu

6.Using the numeric keypad, enter the values for Start, Stop, Tolerance, Dwell time, Max time, number of pressure Points up, and number of pressure Points down, pressing ENTER after each value. The rotary knob may be used to skip fields.

7.Press Program [F1]. The program will be generated and the display will show the first step.

8.Press the Name [F6] key to edit the name of the program. See next section, Chapter 4, Changing the Name of a Program, for creating a user-defined program name.

4-20

Image 44
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii How to Contact Fluke General InformationIntroduction General InformationSafety Information Symbols Used in this Manual FeaturesPower-On Self Test Automatic Zero AdjustEase of Operation Easily ProgrammableStandard Equipment & Options Ruska 7350 Options List Part NumberRuska Power Supply Theory of OperationElectronics Section Back Plane BoardRuska 7350 Block Diagram Motor Controller IEEE-488 InterfaceFront Panel Pneumatic Module 76xx Interface BoardPneumatic Pump GearmotorElectronics Regulator Dome RegulatorSystem Transducer Test PortRuska Installation Unpacking the RuskaDrive Air Supply Port Powering Up the RuskaPneumatic Connections Local Operation Local OperationNumeric Keypad Function KeysCancel, Previous Menu TreeTutorial Main MenuMain Menu Entering Pressure Selecting Language Selecting Mode of OperationTare Mode Setting the Pressure Set-Point Main MenuControlling Pressure Triple Range OptionalVent Step/JogMenu Setup MenuMenu Setup Limits High LimitLow Limit Auto VentControl Band Access Menu Setup UserSlew Rate Slew LimitStep Size Bar Graph MaximumReady Tolerance Gas Head Pressure CorrectionMenu Setup Units Pressure FilterChanging the Number of Decimals Key Click13. Menu Setup Units Edit Name Menu Menu Setup Remote Menu Setup SystemGpib Address ProtocolReset Calibration PasswordDate/Time Menu CalibrateMenu Program Storing a Sequence in Memory Preparing to ProgramEntering a New Program Mode of EntryMaxTime 17. Menu Program Edit Menu Automatically Generating a Program 18. Menu Program Edit Auto MenuChanging the Name of a Program Changing an Existing ProgramChanging the Configuration Stored with a Program 20. Menu Program Edit Menu21. Menu Program Config Menu Running a ProgramMenu Test Menu Test Sweep TestMenu Test Self Test Menu Test Remote TestMenu Test Shop1 26. Menu Test Shop1 MenuMenu Display Menu Display BlankRuska Remote Operation Remote OperationCapabilities IEEE-488Remote/Local Operation RS-232Configuration Scpi Command FormatDevice Messages Scpi Response FormatANSI/IEEE 488.2-1987 Command Summary Scpi Command SummaryMODE? Example Scpi Commands ListScpi Status Registers Interface Panel Emulation Serial Operation Ruska Maintenance Preventive MaintenanceObserving the Ruska 7350’s Full Scale Rating Observing the Software Version NumberProcessor Battery CalibrationCalibration Instructions Single Sensor PreparationStoring the Coefficients Editing the Calibration Coefficients ZeroingFuse CleaningRuska 7350 Pump Lubrication Packing Instructions Preparation for Storage and ShipmentDisconnection Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Shipping Instructions Packing the Calibrator7350 Summary of Specifications AccuracySpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range