| Remote Operation | 5 |
|
| Device Messages | |
:MODE? |
| calibration edit enabled? |
|
:MODE ON¦OFF¦1¦0 |
| enable calibration edit |
|
DISP |
| (Cal. button required) |
|
|
|
| |
:ENABle ON¦OFF¦1¦0 | turns front panel display on/off |
| |
:TEXT <string> |
| displays message on front panel |
|
:BGRaph <number> |
| sets bar graph maximum |
|
OUTPut |
|
|
|
:STATe ON¦OFF¦1¦0 |
| off=MEASure, on=CONTrol |
|
:STATe? |
| returns 0=Measure or 1=Control |
|
:MODE MEASure¦CONTrol¦VENT | sets mode |
| |
:MODE? |
| returns mode string |
|
PROGram |
|
|
|
:CATalog? |
| returns list of defined programs |
|
[SELected] |
|
|
|
:DEFine <program block> | Define program |
| |
|
| press1, toler1, dwell1, max1, press2, |
|
|
| toler2, .. |
|
:DEFine? |
| read program definition |
|
:DELete |
|
|
|
[:SELected] |
| deletes current program |
|
:ALL |
| deletes all programs |
|
:NAME <program name> | select current program |
| |
:STATe RUNPAUSeSTOPCONTinue | set program state |
| |
:STATe? | ` | read program state |
|
:CONFigure |
|
|
|
:RECall |
| restore saved configuration |
|
:SAVE |
| save current configuration |
|
SENSE |
|
|
|
[:PRESSure] |
|
|
|
[:RESolution] <number> | set pressure display resolution |
| |
:AUTO <boolean>ONCE | return to default resolution |
| |
:MODE? |
| returns ABSOLUTE or TARE |
|
:RANGE |
|
|
|
[:UPPer] |
| returns RUSKA 7350 full scale value in |
|
units |
|
|
|
:LOWer |
| returns 0 |
|
:REFerence |
|
|
|
[:HEIGht] <number> | set head height |
| |
:SGRavity <numbers> | set specific gravity |
| |
[SOURCE] |
|
|
|
[:PRESsure] |
|
|
|
[:LEVel] |
|
|
|
[:IMMediate] |
|
|
|
[:AMPLitude] <number> | sets pressure setpoint |
| |
[:AMPLitude]? | read pressure setpoint |
| |
:MODE FIXedLIST | set source parameter set |
| |
:TOLerance <number> | specifies output tolerance |
| |
:SLEW <number> |
| set slew rate |
|
:CONTrol <number> | set control band |
|