Fluke 7350 specifications Setpoint is out of range

Page 78

RUSKA 7350

Users Manual

Value

Description and Corrective Action

 

 

-286

Program Runtime Error. An error occurred while running the program. Usually the

 

setpoint is out of range.

 

 

-313

Calibration Data Lost. The calibration data has been lost and the unit must be

 

recalibrated.

 

 

-315

Configuration Data Lost. The configuration data has been lost. Check all parameters

 

to be sure that they are correct.

 

 

-330

Self Test failed. See display for more information.

 

 

-350

Queue Overflow. The error queue was full and messages were lost.

 

 

-400

Query Error. A read request was received when there was nothing to read.

 

 

500

Controller Malfunction. Internal control failure.

 

 

501

High Limit Exceeded. The pressure was greater than the high limit.

 

 

502

Low Limit Exceeded. The pressure was less than the low limit.

 

 

503

Slew Limit Exceeded. The pressure changed faster than the slew limit allowed.

 

 

521

Pressure Overrange. The pressure reading is outside the range of the RUSKA 7350.

 

 

538

Auto vent. The Vent Limit was exceeded and the instrument entered Vent Mode.

 

 

543

Supply Pressure. The instrument is attempting to increase pressure and the

 

pressure is not responding. Check main pressure supply, shop air supply, and for

 

leaks.

 

 

600

Factory Data Lost. Internal factory constants have been lost. Contact Fluke for more

 

information.

 

 

602

Model Changed. Firmware has been downloaded that is not correct. Download the

 

correct firmware.

 

 

706

Pump position failure. The position sensor is not detecting movement in the pump

 

after the pump has been commanded to move.

 

 

707

Range Switch. The current range can only be changed when the pressure reading is

 

below 100 psi (700 kPa).

 

 

B-2

Image 78
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii Introduction How to Contact FlukeGeneral Information General InformationSafety Information Symbols Used in this Manual FeaturesEase of Operation Power-On Self TestAutomatic Zero Adjust Easily ProgrammableStandard Equipment & Options Ruska 7350 Options List Part NumberRuska Electronics Section Power SupplyTheory of Operation Back Plane BoardRuska 7350 Block Diagram IEEE-488 Interface Front PanelMotor Controller Pneumatic Pump Pneumatic Module76xx Interface Board GearmotorSystem Transducer Electronics RegulatorDome Regulator Test PortRuska Installation Unpacking the RuskaPowering Up the Ruska Pneumatic ConnectionsDrive Air Supply Port Numeric Keypad Local OperationLocal Operation Function KeysCancel, Previous Menu TreeTutorial Main MenuMain Menu Entering Pressure Selecting Mode of Operation Tare ModeSelecting Language Controlling Pressure Setting the Pressure Set-PointMain Menu Triple Range OptionalVent Step/JogMenu Setup MenuLow Limit Menu Setup LimitsHigh Limit Auto VentControl Band Slew Rate AccessMenu Setup User Slew LimitReady Tolerance Step SizeBar Graph Maximum Gas Head Pressure CorrectionChanging the Number of Decimals Menu Setup UnitsPressure Filter Key Click13. Menu Setup Units Edit Name Menu Gpib Address Menu Setup RemoteMenu Setup System ProtocolDate/Time ResetCalibration Password Menu CalibrateMenu Program Storing a Sequence in Memory Preparing to ProgramMode of Entry MaxTimeEntering a New Program 17. Menu Program Edit Menu Automatically Generating a Program 18. Menu Program Edit Auto MenuChanging the Name of a Program Changing an Existing ProgramChanging the Configuration Stored with a Program 20. Menu Program Edit Menu21. Menu Program Config Menu Running a ProgramMenu Test Menu Test Sweep TestMenu Test Self Test Menu Test Remote TestMenu Test Shop1 26. Menu Test Shop1 MenuMenu Display Menu Display BlankRuska Capabilities Remote OperationRemote Operation IEEE-488Remote/Local Operation RS-232Device Messages ConfigurationScpi Command Format Scpi Response FormatANSI/IEEE 488.2-1987 Command Summary Scpi Command SummaryMODE? Example Scpi Commands ListScpi Status Registers Interface Panel Emulation Serial Operation Ruska Observing the Ruska 7350’s Full Scale Rating MaintenancePreventive Maintenance Observing the Software Version NumberProcessor Battery CalibrationCalibration Instructions Single Sensor PreparationStoring the Coefficients Editing the Calibration Coefficients ZeroingCleaning Ruska 7350 Pump LubricationFuse Preparation for Storage and Shipment Disconnection InstructionsPacking Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Shipping Instructions Packing the Calibrator7350 Summary of Specifications AccuracySpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range