4.Press Run [F2]. The configuration of the Calibrator stored with the program is restored, the pressure
5.To pause the program, press Pause [F3]. “Pause” will now be highlighted and the Calibrator will continue controlling to the current
6.To stop the program, press Stop [F5]. The program will stop running but the Calibrator will continue controlling to the current
Menu Test
The Menu Test section of the system is used to perform a number of the diagnostic and tuning functions that are available with the RUSKA 7350.
gld25.bmp
Figure 4-23. Menu Test - Menu
Menu Test — Sweep Test
The sweep function can be used to automatically exercise the elastic sensing element of the device under test prior to performing a calibration. For the Calibrator to perform the sweep test, input the high and low
1.Ensure that the units, limits, and control parameters are set to their desired values.
2.The sweep test is entered from the Sweep Test Menu. From the Main Menu, (press PREVIOUS until the Main Menu appears), press Menu [F6], then Test [F5], and then Sweep [F2].
3.Enter the high and low pressure points, the control tolerance, dwell time, and the number of cycles.