[SENSe:]DETector:BANDwidth {3|20|200|MINimum|MAXimum}
Specify the lowest frequency expected in the input signal. The multimeter
selects the slow, medium (default), or fast ac filter based on the frequency
you specify. MIN = 3 Hz. MAX = 200 Hz. [Stored in volatile memory]
[SENSe:]DETector:BANDwidth? [MINimum|MAXimum]
Query the ac filter. Returns “+3.000000E+00”, “+2.000000E+01”, or
“+2.000000E+02”.
[SENSe:]ZERO:AUTO {OFF|ONCE|ON}
Disable or enable (default) the autozero mode. The OFF and ONCE
parameters have a similar effect. Autozero OFF does not issu e a new
zero measurement until the next time the multimeter goes to the
“wait-for-trigger” state. Autozero ONCE issues an immediate zero
measurement. [Stored in volatile memory]
[SENSe:]ZERO:AUTO?
Query the autozero mode. Returns “0” (OFF or ONCE) or “1” (ON).
INPut:IMPedance:AUTO {OFF|ON}
Disable or enable the automatic input resistance mode for dc voltage
measurements. With AUTO OFF (default), the input resistance is fixed
at 10 MΩ for all ranges. With AUTO ON, the input resistance is set to
>10 GΩ for the 100 mV, 1 V, and 10 V ranges. [Stored in volatile memory]
INPut:IMPedance:AUTO?
Query the input resistance mode. Returns “0” (OFF) or “1” ( ON).
ROUTe:TERMinals?
Query the multimeter to determine if the front or rear input terminals
are selected. Returns “FRON” or “REAR”.