Autozero (continued) The following table shows the relationship between integration time andautozero settings from the remote interface and the correspondingfront-panel settings.
From 34401A Specifications:1
Remote Configuration
NPLC: 100
Autozero: On
Digits Displayed: 61⁄2
NPLC: 100
Autozero: Off
Digits Displayed: 61⁄2
NPLC: 10
Autozero: On
Digits Displayed: 61⁄2
NPLC: 10
Autozero: Off
Digits Displayed: 61⁄2
NPLC: 1
Autozero: On
Digits Displayed: 51⁄2
NPLC: 1
Autozero: Off
Digits Displayed: 51⁄2
NPLC: 0.2
Autozero: On
Digits Displayed: 51⁄2
NPLC: 0.2
Autozero: Off
Digits Displayed: 51⁄2
NPLC: 0.02
Autozero: On
Digits Displayed: 41⁄2
NPLC: 0.02
Autozero: Off
Digits Displayed: 41⁄2
Front-Panel Equivalent
Slow 6 digits
N/A
Fast 6 digits
Slow 5 digits
N/A
Slow 4 digits
N/A
N/A
Fast 5 digits
N/A
Fast 4 digits
Digits Displayed
N/A
6
1⁄2
N/A
6
1⁄2
N/A
5
1⁄2
N/A
5
1⁄2
N/A
4
1⁄2
Readings/Sec
N/A
0.6
N/A
6
N/A
60
N/A
300
N/A
1000
1 See the Agilent 34401A specif ications listed on page 21 7.
Chapter 3 Features and Functions Measurement Configuration 60