Autozero (continued) The following table shows the relationship between integration time andautozero settings from the remote interface and the correspondingfront-panel settings.
From 34401A Specifications:1

Remote Configuration

NPLC: 100
Autozero: On
Digits Displayed: 612
NPLC: 100
Autozero: Off
Digits Displayed: 612
NPLC: 10
Autozero: On
Digits Displayed: 612
NPLC: 10
Autozero: Off
Digits Displayed: 612
NPLC: 1
Autozero: On
Digits Displayed: 512
NPLC: 1
Autozero: Off
Digits Displayed: 512
NPLC: 0.2
Autozero: On
Digits Displayed: 512
NPLC: 0.2
Autozero: Off
Digits Displayed: 512
NPLC: 0.02
Autozero: On
Digits Displayed: 412
NPLC: 0.02
Autozero: Off
Digits Displayed: 412

Front-Panel Equivalent

Slow 6 digits
N/A
Fast 6 digits
Slow 5 digits
N/A
Slow 4 digits
N/A
N/A
Fast 5 digits
N/A
Fast 4 digits

Digits Displayed

N/A
6
12
N/A
6
12
N/A
5
12
N/A
5
12
N/A
4
12

Readings/Sec

N/A
0.6
N/A
6
N/A
60
N/A
300
N/A
1000
1 See the Agilent 34401A specif ications listed on page 21 7.
Chapter 3 Features and Functions Measurement Configuration 60