Integration Time
(continued)
Front-Panel Operation: Integration time is set indirectly when you
select the number of digits. See the table f or resolution on page 54.
Remote Interface Operation:
<function>:NPLCycles {0.02|0.2|1|10|100|MINimum|MAXimum}
For frequency and period measurements, aperture time (or gate time)
is analogous to integration time. Specify 10 ms (412 digits), 100 ms
(default; 512 digits), or 1 second (612 digits).
FREQuency:APERture {0.01|0.1|1|MINimum|MAXimum}
PERiod:APERture {0.01|0.1|1|MINimum|MAXimum}
Front / Rear Input Terminal Switching
Any measurement made using the front terminals can also be made
using the input terminals on the rear panel. See “The Front Panel at
a Glance,” on page 2, for the location of the front / rear switch.
The input terminals can only be configured from the fr ont panel.
You cannot select the terminals from the remote interface, but you
can query the present setting.
The Rear annunciator turns on w hen you select the rear termi nals.
Remote Interface Operation: You can query the multimeter to
determine whether the front or rear input terminals are selected.
ROUTe:TERMinals? returns “FRON” or “REAR
Chapter 3 Features and Functions
Measurement Configuration