The Standard Event Regist er
The standard event register reports the following types o f instrument
events: power-on detected, command syntax errors, comma n d ex ec utio n
errors, self-test or calibration errors, que ry errors, or when an *OPC
command is executed. Any or all of these conditions can be repor t ed in
the standard event summary bit through the enable register. You must
write a decimal value using the *ESE (event status enable) command to
set the enable register mask.
An error condition (standard event regi ster bits 2, 3, 4, or 5) will always
record one or more errors in the multi meter’s error queue, except for the
following case. Read the error queue using SYSTem:ERRor?.
A reading overload condi tion is al ways re ported in both the standa rd event
register (bit 3) and th e questio nable data ev ent reg ister ( bits 0, 1 , or 9).
However, no error message is recorded in the multim et er’s error queue.
Bit Definitions – Standard Event Register
Bit
0 Operation Complete
1 Not Used
2 Query Error
3 Device Error
4 Execution Error
5 Command Error
6 Not Used
7 Power On
Decimal
Value
1
2
4
8
16
32
64
128
Definition
All commands prior to and including an *OPC
command have been executed.
Always set to 0.
The multimeter tried to read the output buffer but
it w as empty. Or, a new com mand line was
received before a previous query has been read.
Or, both the input and outpu t buffers are full.
A self-test, calibration, or reading overload error
occurred (see error numbers 501 through 748 in
chapter 5).
An execution error occurred (see error numbers
-211 t hrough -230 in chapter 5).
A command syntax error occurred (see error
numbers -101 through -158 in chapter 5).
Always set to 0.
Power has been turned off and on since the last
time the event register was read or cleared.
Chapter 4 Remote Interface Reference
The SCPI Status Model
140