Automatic Trigger Delays

If you do not specify a trigger delay, the multimeter selects an

automatic delay for you. The delay is determined by function, range,

integration time, and ac filter setting.

DC Voltage and DC Current (for all ranges):
Resistance (2-wire and 4-wire):
AC Voltage and AC Current (for all ranges):
Frequency and Period:
Integra tion Time
NPLC 1
NPLC < 1
Trigger Delay
1.5 ms
1.0 ms
Range
100
1 k
10 k
100 k
1 M
10 M
100 M
Trigger Delay
(For NPLC 1)
1.5 ms
1.5 ms
1.5 ms
1.5 ms
15 ms
100 ms
100 ms
AC Filter
Slow
Medium
Fast
Trigger Delay
7.0 sec
1.0 sec
600 ms
Remote or single/external trigge r
Range
100
1 k
10 k
100 k
1 M
10 M
100 M
Trigger Delay
(For NPLC < 1)
1.0 ms
1.0 ms
1.0 ms
1.0 ms
10 ms
100 ms
100 ms
AC Filter
Slow
Medium
Fast
Trigger Delay
1.5 sec
200 ms
100 ms
Front panel with auto trigger ON
Trigger De lay
1.0 sec
Remote or single/external trigger
Trigger De lay
0 sec
Front panel with auto trig ger ON
3

Chapter 3 Features and Functions

Triggering

81