Chapter 8 Specifications

DC Characteristics

DC Characteristics

 

 

 

 

 

Accuracy Specifications ± ( % of reading + % of range )

[ 1 ]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temperature

 

 

 

 

Test Current or

24 Hour [ 2 ]

90 Day

1 Year

Coefficient /°C

 

Function

Range [ 3 ]

Burden Voltage

23°C ± 1°C

23°C ± 5°C

23°C ± 5°C

0°C – 18°C

 

 

 

 

 

 

 

 

 

28°C – 55°C

 

 

 

 

 

 

 

 

 

 

DC Voltage

100.0000 mV

 

0.0030 + 0.0030

0.0040 + 0.0035

0.0050 + 0.0035

0.0005

+ 0.0005

 

 

1.000000

V

 

0.0020 + 0.0006

0.0030 + 0.0007

0.0040 + 0.0007

0.0005

+ 0.0001

 

 

10.00000

V

 

0.0015 + 0.0004

0.0020 + 0.0005

0.0035 + 0.0005

0.0005

+ 0.0001

 

 

100.0000

V

 

0.0020 + 0.0006

0.0035 + 0.0006

0.0045 + 0.0006

0.0005

+ 0.0001

 

 

1000.000

V

 

0.0020 + 0.0006

0.0035 + 0.0010

0.0045 + 0.0010

0.0005

+ 0.0001

 

 

 

 

 

 

 

 

 

 

Resistance

100.0000

1 mA

0.0030 + 0.0030

0.008 + 0.004

0.010 + 0.004

0.0006

+ 0.0005

[ 4 ]

1.000000 k

1 mA

0.0020 + 0.0005

0.008 + 0.001

0.010 + 0.001

0.0006

+ 0.0001

 

 

10.00000 k

100 A

0.0020 + 0.0005

0.008 + 0.001

0.010 + 0.001

0.0006

+ 0.0001

 

 

100.0000 k

10 A

0.0020 + 0.0005

0.008 + 0.001

0.010 + 0.001

0.0006

+ 0.0001

 

 

1.000000

M

5 A

0.002 + 0.001

0.008 + 0.001

0.010 + 0.001

0.0010

+ 0.0002

 

 

10.00000

M

500 nA

0.015 + 0.001

0.020 + 0.001

0.040 + 0.001

0.0030

+ 0.0004

 

 

100.0000

M

500 nA 10 M

0.300 + 0.010

0.800 + 0.010

0.800 + 0.010

0.1500

+ 0.0002

 

 

 

 

 

 

 

 

 

 

 

DC Current

10.00000 mA

< 0.1 V

0.005

+ 0.010

0.030 + 0.020

0.050 + 0.020

0.002

+ 0.0020

 

 

100.0000 mA

< 0.6 V

0.01

+ 0.004

0.030 + 0.005

0.050 + 0.005

0.002

+ 0.0005

 

 

1.000000

A

< 1 V

0.05

+ 0.006

0.080 + 0.010

0.100 + 0.010

0.005

+ 0.0010

 

 

3.000000

A

< 2 V

0.10

+ 0.020

0.120 + 0.020

0.120 + 0.020

0.005

+ 0.0020

 

 

 

 

 

 

 

 

 

 

 

Continuity

1000.0

 

1 mA

0.002

+ 0.030

0.008 + 0.030

0.010 + 0.030

0.001 + 0.002

 

 

 

 

 

 

 

 

 

 

 

Diode Test

1.0000 V

 

1 mA

0.002

+ 0.010

0.008 + 0.020

0.010 + 0.020

0.001 + 0.002

[12]

 

 

 

 

 

 

 

 

 

 

DC:DC Ratio

100 mV

 

 

( Input Accuracy ) + ( Reference Accuracy )

 

 

 

 

to

 

 

 

 

 

 

 

 

 

 

1000 V

 

 

Input Accuracy = accuracy specification for the HI-LO input signal.

 

 

 

 

 

 

Reference Accuracy = accuracy specification for the HI-LO reference input signal.

 

 

 

 

 

 

 

 

 

 

 

Transfer Accuracy ( typical )

(24 hour % of range error )

2

Conditions:

￿Within 10 minutes and ± 0.5°C.

￿Within ±10% of initial value.

￿Following a 2-hour warm-up.

￿Fixed range between 10% and 100% of full scale.

￿Using 612 digit slow resolution ( 100 PLC ).

￿Measurements are made using accepted metrology practices.

216

Page 222
Image 222
Agilent Technologies 34401A manual DC Characteristics

34401A specifications

Agilent Technologies 34401A is a highly reputable digital multimeter (DMM), renowned for its accuracy, versatility, and performance in a wide range of measurement applications. First introduced in the early 1990s, the 34401A has become a staple in laboratories, manufacturing environments, and educational institutions due to its robust build and comprehensive features.

One of the standout characteristics of the 34401A is its impressive accuracy, boasting a basic DC voltage accuracy of 0.0035%, making it ideal for precise measurements in electronic testing. The multimeter offers a wide range of measurement capabilities, including voltage (DC and AC), current (DC and AC), resistance, frequency, and continuity test, making it a highly versatile tool for engineers and technicians.

The 34401A utilizes a 6½-digit resolution, enabling the user to measure small changes in electrical signals with remarkable precision. With a sampling rate of up to 1000 readings per second, it can handle dynamic test requirements, providing timely results without sacrificing measurement integrity.

In terms of connectivity, the Agilent 34401A includes a GPIB (General Purpose Interface Bus) for remote operation, allowing users to automate their testing processes and integrate the multimeter easily into larger measurement systems. This feature is particularly beneficial in automated test environments, where speed and accuracy are paramount.

Additionally, the DMM incorporates advanced measurement functions, such as statistical analysis, including mean, standard deviation, and minimum/maximum readings, which provide users with valuable insights into their data. The instrument's user-friendly interface, complete with a clear digital display and intuitive controls, enhances usability and ensures that users can easily navigate its various functions.

Robust software support further extends the capabilities of the Agilent 34401A. It is compatible with a range of software tools for data logging and analysis, enabling users to efficiently document and analyze their measurements over time.

In summary, Agilent Technologies 34401A stands out in the realm of digital multimeters due to its high accuracy, extensive measurement functionalities, and user-friendly features. Whether for research and development, quality control, or educational purposes, the 34401A remains a preferred choice for professionals demanding reliable and precise measurement solutions. Its enduring legacy in the industry speaks to its performance and reliability, making it an essential tool for anyone involved in electronic testing and measurement.