4.Set Transient Level-The transient resistance level "R:TLV" is meaningful only if transient operation (described later) is turned on.

a. Set the transient level to 0.8 ohm by pressing

. Remember that in the low range the transient level must be set higher than the main level.

b.Press again and note that the display indicates "R:TLV 0.8000". Note that you can use the Input ENTRY keys to increment and decrement the transient resistance level. Operation is similar to that described for the main resistance level.

Setting CV Values

The CV values for the are programmed by pressing the applicable FUNCTION keys and setting the desired values using the ENTRY keys. The display identifies the selected function; for example "V:TLV" identifies the transient voltage level.

Range

The voltage values can only be programmed in one range. The valid CV values are listed in Table 4-4 along with the applicable front panel key and display identifier. All voltage levels are programmed in volts and the voltage slew rate is programmed in volts/microsecond.

Function

Set Main Level Set Slew Rate

Set Transient Level

Notes:

 

Table 4-4. CV Programming Ranges

 

 

Key

Display

Range of Values

 

 

6060B

 

6063B

 

"VOLT value"

0.000 to 60.000

 

0.000 to 240.000

 

"V:SLW value"

0.0010 to 0.5000 (V/ μs)

(Note 1)

0.0040 to 2.000 (V/ μs)

 

(shifted)

 

 

 

 

 

"V:TLV value"

0.000 to 60.000

(Note 2)

0.000 to 240.00

1.There are 12-discrete steps within the voltage slew range. Because of bandwidth limitations, only 9 slew rate steps can be achieved (see Table 1-1). Any slew rate can be programmed. (There are no upper and lower limits that would cause an error.) The Electronic Load automatically selects one of the 12 slew rates that is closest to the programmed value. See Chapter 2 Operation Overview in this manual.

2.The transient voltage level is meaningful only if transient operation is turned on. The transient voltage level must be set to a higher value than the main voltage level. See Transient Operation.

Examples

The following examples illustrate how to program CV values. Before you do these examples, press to set the CV values to their factory default values.

1.Set Main level

a.Set the main voltage level to 20 volts by pressing

b.Press again and check that the display indicates "VOLT 20.000".

Local Operation 59

Page 59
Image 59
Agilent Technologies 5951-2827, 5951-2826 manual Setting CV Values, Range