Program Listing (continued)

520OUTPUT @Ld;"*TRG"

530IF Flag THEN WAIT 30

540INPUT "Enter current through shunt for high point in amps",Trpt_curr

550OUTPUT @Ld;"CAL:TLEV";Trpt_curr

560OUTPUT @Ld;"TRAN OFF"

570PRINT "Test unit to verify that transient values are in spec"

580PRINT "Press CONT when ready to calibrate next range or mode"

590PAUSE

600SUBEND

610!

620SUB Cal_volt(@Ld,Chan,Volt_hipt,Volt_lopt)

630PRINT "VOLTAGE CALIBRATION"

640PRINT "Set power supply according to calibration information table"

650PRINT "Press CONT when ready"

660PAUSE

670OUTPUT @Ld;"CHAN";Chan

680OUTPUT @Ld;"MODE:VOLT"

690OUTPUT @Ld;"VOLT";.05*Volt_hipt

700

WAIT 3

710INPUT "Enter voltage across input terminals for low point in volts",Lopt_v

720OUTPUT @Ld;"CAL:LEV:LOW";Lopt_volts

730OUTPUT @Ld;"CAL:MEAS:LOW";Lopt_volts

740OUTPUT @Ld;"VOLT";.85*Volt_hipt

750

WAIT 3

760INPUT "Enter voltage across input terminals for high point in volts", Hipt_

770OUTPUT @Ld;"CAL:LEV:HIGH";Hipt_volts

780OUTPUT @Ld;"CAL:MEAS:HIGH";Hipt_volts

790OUTPUT @Ld;"VOLT";Volt_lopt

800

WAIT 3

810INPUT "Enter voltage across input terminals for low point in volts",Lopt_v

820OUTPUT @Ld;"CAL:LEV:LOW";Lopt_volts

830OUTPUT @Ld;"CAL:MEAS:LOW";Lopt_volts

840OUTPUT @Ld;"VOLT";Volt_hipt

850

WAIT 3

860INPUT "Enter voltage across input terminals for high point in volts", Hipt_

870OUTPUT @Ld;"CAL:LEV:HIGH";Hipt_volts

880OUTPUT @Ld;"CAL:MEAS:HIGH";Hipt_volts

890PRINT "Test unit to verify that program and readback values are in spec"

900PRINT "Press CONT when ready to calibrate transient level"

910PAUSE

920OUTPUT @Ld;"VOLT";Volt_lopt

930OUTPUT @Ld;"TRAN:STAT ON;MODE TOGG;:TRIG:SOUR BUS"

940OUTPUT @Ld;"VOLT:TLEV";Volt_hipt

950OUTPUT @Ld;"*TRG"

960INPUT "Enter voltage across input terminals for transient point in volts"

970OUTPUT @Ld;"CAL:TLEV";Trpt_volts

980OUTPUT @Ld;"TRAN OFF"

990PRINT "test unit to verify that transient values are in spec"

1000 PRINT "Press CONT when ready to calibrate next mode"

1010 PAUSE

1020 SUBEND