Table 5-1. Remote Programming Ranges (continued)

Function

HPSL Command

(Short Form)

Range of Values

Constant Resistance (CR)

Set Range

Low Range

Middle Range

High Range

Set Main Level

Low Range

Middle Range

High Range

Set Slew Rate

"RES:RANG value"

’RES value"

6060B

³0 or £ 1 W >1 W and £ kW >1 kW and £ kW

0 to 1 W

1W to 1 kW

10W to 10 kW

6063B

³0 and £ 24 W

£24 W and £ 24 kW >24 kW and £ 240 kW

0 to 24 W

24W to 24 kW

240W to 240 kW

Low Range

Middle/High Range

Set Transient Level

*Set Triggered Level

Constant Voltage (CV)

Set Main Level

Set Slew Rate

Set Transient Level

*Set Triggered Level

Transient Operation

Set Frequency

Set Duty Cycle

*Set Pulse Width

"VOLT: SLEW value" "CURR:SLEW value"

"RES:TLEV value" "RES:TRIG value"

"VOLT value"

"VOLT:SLEW value" "VOLT:TLEV value" "VOLT:TRIG value"

"TRAN:FREQ value" "TRAN:DCYC value"

"TRAN:TWID value"

Same as CV slew rate

Same as CC slew rate

Same as main CR level

Same as main CR level

6060B

6063B

0 to 60 V

0 to 240 V

1000 to 5,000,000 V/s

4000 to 2,000,000 V/s

Same as main CV level

Same as main CV level

0.25 to 10000 Hz

3 to 97% (0.25 Hz to 1 kHz

6-94% (1 kHz-10 kHz)

0.00005 to 4 seconds

Current Protection

*Set Current Level

*Set Delay Time

 

6060B

6063B

"CURR:PROT value"

0 to 61.2 A

0 to 10.2 A

"CURR:PROT:DEL value"

 

0 to 60 seconds

*Can only be programmed remotely via the GPIB.

Remote Operation 73