Explanation

 

LINE 10-20

Specify select code, address, and channel (default 705, 1)

LINE 30

Turn calibration mode on

LINE 40-90

Assign variables for subprograms (see module calibration tables)

LINE 100

Store new constants in EEROM when calibration complete

LINE 110

Turn calibration mode off

LINE 140

Current calibration subroutine

LINE 200-220

Select channel, current mode, and range

LINE 230

Set high calibration point

LINE 240

If high current range, wait for internal current shunt to stabilize

LINE 260

Send measurement in amperes for high main calibration point

LINE 270

If high current range, send measurement in amperes for high readback cal point

LINE 280

Set low calibration point

LINE 300

Send measurement in amperes for low main calibration point

LINE 310

If high current range, send measurement in amperes for low readback cal point

LINE 350

Set low calibration point

LINE 360-370

Select transient toggle mode and GPIB trigger source

LINE 380-390

Turn transient mode on and set transient calibration point

LINE 400

Trigger transient level

LINE 410

If high current range, wait for internal current shunt to stabilize

LINE 430

Send measurement in amperes for high transient calibration point

LINE 440

Turn transient mode off

LINE 500

Voltage calibration subroutine

LINE 550-560

Select channel and voltage mode

LINE 570

Set high calibration point

LINE 590

Send measurement in volts for high main calibration point

LINE 600

Send measurement in volts for high readback calibration point

LINE 610

Set low calibration point

LINE 630

Send measurement in volts for low main calibration point

LINE 640

Send measurement in volts for low readback calibration point

LINE 680

Set low calibration point

LINE 690-700

Select transient toggle mode and GPIB trigger source

LINE 710-720

Turn transient mode on and set transient calibration point

LINE 730

Trigger transient level

LINE 750

Send measurement in volts for transient calibration point

LINE 760

Turn transient mode off

LINE 820

Resistance calibration subroutine

LINE 870-890

Select channel, resistance mode, and range

LINE 900

Set high calibration point

LINE 930-940

Calculate and send measurement in ohms for high main calibration point

LINE 950

Set low calibration point

LINE 980-990

Calculate and send measurement in ohms for low main calibration point

LINE 1030-1070

If middle and high range, set high calibration point; otherwise set low point

LINE 1080-1090

Select transient toggle mode and GPIB trigger source

LINE 1100

Turn transient mode on

LINE 1110-1150

If middle and high range, set lower transient point; otherwise set higher point

LINE 1160

Trigger transient level

LINE 1190-1200

Calculate and send measurement in ohms for transient calibration point

LINE 1210

Turn transient mode off

Calibration 85

Page 85
Image 85
Agilent Technologies 5951-2827, 5951-2826 manual Explanation, Line