Short On/Off
A load can simulate a short circuit at its input by turning the load on with
on/off at the front panel ( key) or via the GPIB (INPUT:SHORT ONOFF command). The short on/off change uses the slew rate setting of the active mode and range.
Figure 2-8. Transition Times and Slew Rates
The actual value of the electronic short is dependent on the mode and range that are active when the short is turned on. In CV mode, it is equivalent to programming zero volts. In CC mode, it is equivalent to programming
Note that turning the short on in CV mode may cause the load to draw so much current that the software current limit operates, which may turn the input off.
Turning the short circuit on does not affect the programmed settings, and the input will return to the previously programmed values when the short is turned off.
Pressing the Short on/off key with certain user applications may cause damage to the equipment being tested, which may result in personal injury. Contact your Agilent Sales and Service office if you need to have the Short on/off key disabled.