Table 4-6. Factory Default Settings
Function |
| Setting |
| 6060B | 6063B |
Input on/off | on | on |
Short on/off | off | off |
CURR level | 0 A | 0 A |
CURR transient level | 0 A | 0 A |
CURR slew rate | 1 A/μs | 0.83 A/μs |
CURR range | 60 A | 10 A |
*CURR protection level | 61.2 A | 10.2 A |
*CURR protection delay | 15 s | 15 s |
*CURR protection on/off | off | off |
VOLT level | 60 V | 240 V |
VOLT transient level | 60 V | 240 V |
VOLT slew rate | 5 V/μs | 2 V/μs |
RES level | 1000 Ω | 2400 Ω |
RES transient level | 1000 Ω | 24000 Ω |
RES range | 1000 Ω | 24000 Ω |
Mode | CC | CC |
Transient Operation | off | off |
Frequency | 1 kHz | 1 kHz |
Duty Cycle | 50% | 50% |
**Transient mode | continuous | continuous |
*Pulse width | 0.5 ms | 0.5 ms |
*Port0 output signal | TTL logic 0 | TTL logic 0 |
*Calibration mode | off | off |
*Trigger source | hold | hold |
*Can only be programmed remotely via the GPIB.
**Continuous transient operation is the only mode of transient operation available at the front panel. Pulsed, toggled, and continuous transient operating modes may be programmed remotely via the GPIB.
Changing "Wake-up" Settings
The
You can change the
(blue shift key) | (shifted Recall key) | . |
|
|
When power is turned off and on, the Electronic Load will be set to the values you saved in register 0.
The Save 0 operation takes a few seconds to complete. Do not turn power off until the "SAVE 0 " message goes away indicating that the operation is complete. If you turn off power before completion, the Electronic Load’s non- volatile memory will be corrupted and the Electronic Load
will need to be recalibrated.