Program Listing (continued)

1030

!

1040

SUB Cal_res(@Ld,Chan,Res_rng,Res_hipt,Res_lopt,Flag)

1050

PRINT "RESISTANCE CALIBRATION, RANGE";Res_rng

1060

PRINT "Set power supply to calibration information table"

1070

PRINT "Press CONT when ready to continue"

1080

PAUSE

1090

OUTPUT @Ld;"CHAN";Chan

1100

OUTPUT @Ld;"MODE:RES"

1110

OUTPUT @Ld;"RES:RANG";Res_rng

1120

OUTPUT @Ld;"RES";Res_hipt

1130

INPUT "Enter voltage across input terminals in volts",Hipt_volt

1140

INPUT "Enter current through current shunt in amps”,Hipt_curr

1150

Hipt_res=Hipt_volt/Hipt_curr

1160

OUTPUT @Ld;"CAL:LEV:HIGH";Hipt_res

1170

OUTPUT @Ld;"RES";Res_lopt

1180

INPUT "Enter voltage across input terminals in volts",Lopt_volt

1190

INPUT "Enter current through current shunt in amps",Lopt_curr

1200

Lopt_res=Lopt_volt/Lopt_curr

1210

OUTPUT @Ld;"CAL:LEV:LOW;Lopt_res

1220

PRINT "Test unit to verify resistance values"

1230

PRINT "Press CONT when ready to calibrate transient level"

1240

PAUSE

1250

IF Flag THEN

1260

OUTPUT @Ld;"RES";Res_hipt

1270

ELSE

1280

OUTPUT @Ld;"RES";Res_lopt

1290

END IF

1300

OUTPUT @Ld;"TRAN:STAT ON;MODE TOGG;:TRIG:SOUR BUS"

1310

IF Flag THEN

1320

OUTPUT @Ld;"RES:TLEV";Res_lopt

1330

ELSE

1340

OUTPUT @Ld;"RES:TLEV";Res_hipt

1350

END IF

1360

OUTPUT @Ld;"*TRG"

1370

INPUT "Enter voltage across input terminals in volts",Tran_volt

1380

INPUT "Enter current through current shunt in amps",Tran_curr

1390

Tran_res=Tran_volt/Tran_curr

1400

OUTPUT @Ld;"CAL:TLEV";Tran_res

1410

OUTPUT @Ld;"TRAN OFF"

1420

PRINT "Test unit to verify transient values are in spec"

1430

PRINT "Press CONT when ready to end program or calibrate next range"

1440

PAUSE

1450

SUBEND

84 Calibration

Page 84
Image 84