Table 6-2. Calibration Information

 

 

 

 

 

 

 

6060B

 

 

6063B

 

Ranges and Calibration

Variables

 

Variable

Power

Current

Variable

Power

Current

Points

 

 

Values

Supply

Shunt

Values

Supply

Shunt

 

 

 

 

Settings

 

 

Settings

 

High Current Range

Hi_curr_rng

60

5 V/61 A

100 A

10

25 V/10.5 A

15 A

High Current Offset

Hi_curr_offset

0.0282

 

 

0.0048

 

 

Low Current Range

Lo_curr_rng

6

5 V/10 A

15 A

1

25 V/2 A

15A

Low Current Offset

Lo_curr_offset

0.0197

 

 

0.0032

 

 

Voltage Range

N/A

 

N/A

61 V/5 A

N/A

N/A

246 V/0.6 A

N/A

Voltage Hi point

Volt_hipt

60

 

 

240

 

 

Voltage Lo point

Volt_lopt

2.7

 

 

2

 

 

Low Resistance Range

Lo_res_rng

1

15 V/10.9 A

15 A

24

60 V/1.8 A

15 A

Low Resistance Hi point

Lo_res_hipt

1

 

 

23.9

 

 

Low Resistance Lo point

Lo_res_lopt

.04

 

 

.88

 

 

Middle Resistance Range

Mid_res_rng

10

10.9 V/15 A

15 A

240

43.6 V/4 A

15 A

Middle Resistance Hi point

Mid_res_hipt

30

 

 

500

 

 

Middle Resistance Lo point

Mid_res_lopt

1

 

 

24

 

 

High Resistance Range

Hi_res_rng

1001

60 V/6 A

15 A

24020

*240 V/2 A

15 A

High Resistance Hi point

Hi_res_hipt

120

 

 

2000

 

 

High Resistance Lo point

Hi_res_lopt

12

 

 

240

 

 

78 Calibration

Page 78
Image 78
Agilent Technologies 5951-2826, 5951-2827 manual Calibration Information 6060B 6063B, Supply Shunt Values Settings