80 OUTPUT 705;"TRAN ON
90 OUTPUT 705;"INPUT ON"
| . |
| . |
| . |
| . |
| 200 OUTPUT 705;"*TRG" |
| 210 END |
Line 10: | Turns off Electronic Load input. |
Line 20: | Selects the CR mode. |
Line 30: | Selects the main resistance level to 10 ohms. |
Line 40: | Sets the transient resistance level to 5 ohms. Remember in the 1 to 1k range, the transient resistance level |
| must be set to a lower level than the main resistance level. |
Line 50: | Selects the GPIB as the trigger source. |
Line 60: | Sets the current slew rate to the fastest rate. Remember that in the middle range, the resistance slew rate |
| is programmed in amps/second. |
Line 70: | Selects pulsed transient operation and sets the pulse width to 1 millisecond. |
Line 80: | Turns on the transient generator. |
Line 90 | Turns on Electronic Load input. |
Line 100 |
|
to | Other commands are executed. |
Line 190 |
|
Line 200: The *TRG command generates a 1 millisecond pulse at the Electronic load input.
Table 5-1. Remote Programming Ranges
Function
Constant Current (CC)
Set Range
Low Range
High Range
Set Main Level
Low Range
High Range
Set Slew Rate
Low Range
High Range
Set Transient Level
*Set Triggered Level
HPSL Command
(Short Form)
"CURR:RANG value"
’CURR value"
"CURR:SLEW value"
.
"CURR:TLEV value" "CURR:TRIG value"
Range of Values | ||
6060B | 6063B | |
³ 0 and £ 6 A | ³ 0 and £1 A | |
> 6 and £ 60 A | > 1 and £ 10 A | |
0 to 6 A | 0 to l A | |
0 to 60 A | 0 to 10 A | |
100 to 500,000 A/s | 1.7 to 83,000 A/s | |
1000 to 5,000,000 A/s | 17 to 830,000 A/s |
Same as CC main level
Same as CC main level