Table 2-1.  Test Equipment Required
| TYPE | REQUIRED CHARACTERISTICS | USE | RECOMMENDED MODEL | |
| Oscilloscope | Sensitivity: 1 mV | P,T | Agilent 54504A | |
| 
 | Bandwidth: 20MHz & 100MHz | 
 | 
 | |
| 
 | Input: differential, 50 Ω & 10MΩ | 
 | 
 | |
| RMS Voltmeter | True rms, 10MHz bandwidth | P | Agilent 3400B | |
| 
 | Sensitivity: 1 mV Accuracy: 5% | 
 | 
 | |
| Logic Pulser | 4.5 to 5.5Vdc @ 35mA | T | Agilent 546A | |
| Multimeter | Resolution: 100nV | P,C,T | Agilent 3458A | |
| 
 | Accuracy: 0.0035%, 6½ digit | 
 | 
 | |
| Signature Analyzer | 
 | T | Agilent 5004A | |
| GPIB Controller | Full GPIB capabilities | C,T,P | HP Series, 200/300 | |
| Current Probe | No saturation at 100A | P | Tek P6303 Probe | |
| 
 | Bandwidth: 20Hz to 15MHz | 
 | AM503 Amp | |
| 
 | 
 | 
 | 
 | TM500 Power Module | 
| Electronic Load | 6030A 200 V/5 A | P,C | 6050A & 60503A (4) | |
| 
 | 
 | 60 V/17A | 
 | 6050A & 60503A (4) | 
| 
 | 6031A 20 V/50 A | 
 | 6050A & 60504A (2) | |
| 
 | 
 | 7 V/120 A | 
 | 6050A & 60504A (2) | 
| 
 | 6032A 60 V/17.5 A | 
 | 6050A & 60504A (2) | |
| 
 | 
 | 20 V/50 A | 
 | 6050A & 60504A (2) | 
| 
 | 6035A 500 V/2 A | 
 | Power Resistor 250Ω 1% 1KW | |
| 
 | 
 | 200 V/5 A | 
 | 6050A & 60503A (4) | 
| Current Shunt | 0.1Ω/0.04%/15A/25W | P,C | Guildline 9230/15 | |
| 
 | 0.01Ω/0.04%/100A/100W | 
 | Guildline 9230/100 | |
| 
 | 0.001Ω/0.04%/300A/100W | 
 | Guildline 9230/300 | |
| Power Resistors* | 6030A 3.5Ω/40Ω 1% 1000 W (min.) | P,C | 
 | |
| 
 | 6031A 0.069Ω/0.04Ω 1% 1000W (min.) | 
 | 
 | |
| 
 | 6032A 0.4Ω/3.5Ω 1% 1000W (min.) | 
 | 
 | |
| 
 | 6035A 40Ω/250Ω 1% 1000W (min.) | 
 | 
 | |
| Calibration and Test | Value: | 100Ω, 5%, 1W | C,T | 
 | 
| Resistors | 
 | 1Ω, 5%, ½W | 
 | 
 | 
| 
 | 
 | 1KΩ, 5%, ¼W | 
 | 
 | 
| 
 | 
 | 2KΩ, 0.01%, ¼W | 
 | 
 | 
| Terminating | Value: 50Ω ±5%, noninductive | P | 
 | |
| Resistors (2) | 
 | 
 | 
 | 
 | 
| Blocking | Value: 0.01∝F, 100Vdc | P | 
 | |
| Capacitors (2) | 
 | 
 | 
 | 
 | 
| Variable Voltage | Adjustable from  | P,C | 
 | |
| Transformer | voltage range, 4KVA minimum | 
 | 
 | |
| (autotransformer) | 
 | 
 | 
 | 
 | 
| Isolation Transformer | 4KVA minimum | T | 
 | |
| ≥3.7∝H/turn2 | P | |||
| Toroidal Core | ≅23mm I.D. | 
 | ||
| 
 | 
 | 
 | 
 | Agilent  | 
| DC Power Supply | Voltage range:  | C,T | Agilent 6024A | |
| 
 | Current range:  | 
 | 
 | |
| P = performance | testing | C = calibration adjustments | T = troubleshooting | |
*Not required if using electronic load.
**Less accurate, and less expensive, 
12