Constant Voltage Zero Calibration

Note: Perform this procedure only after completing Voltage Monitor and Remote Readback Full Scale Calibration.

a.Send string "VSET 0; ISET 5; OUT ON".

b.Connect an external supply to the power supply as shown in Figure 2-2.

c.Attach the DVM from - S to + S on the rear panel.

d.Adjust A8R40 (CV PROG ZERO) to 0 ±120∝V.

Current Monitor Zero Calibration

a.Send string "VSET 0; ISET 0; OUT OFF''.

b.Connect a short across power supply output terminals.

c.Attach the DVM from M to IM on the rear panel.

d.Allow several minutes (3 or more) to ensure thermal settling.

e.Adjust A2R8 (I-MON ZERO) to 0 ±100∝V.

Constant Current Zero Calibration

a.Connect the test setup shown in Figure 2-3.

b.Send string ''VSET 5; ISET 0; OUT ON''.

c.Allow several minutes (3 or more) to ensure thermal settling.

d.Adjust A8R29 (CC PROG ZERO) to: 0V ±0.6mV (6030A, 6035A)

0V ±8.0mV (6031A)

0V ±1.7mV (6032A)

Figure 2-3. CC Zero Calibration Setup