Table 2-2. Guide to Recalibration After Repair

Printed Circuit

Block Name

Ref.

Perform These Procedures

 

Board

 

Desig.

 

 

 

 

 

 

A1

Main Board

 

R3

Current Monitor Full Scale Calibration

 

 

 

 

Constant Current Full Scale Calibration

 

 

 

 

 

A1

Main Board

 

T1, T2

Power Limit Calibration

 

 

 

 

 

A5

Diode Board

 

CR4

Power Limit Calibration

 

 

 

 

 

A2

Control Board

Constant Voltage Circuit

All

Voltage Monitor Zero Calibration

 

 

(All Except Current Source)

 

Common Mode Calibration

 

 

 

 

Remote Readback Zero Calibration

 

 

 

 

Constant Voltage Full Scale Calibration

 

 

 

 

Voltage Monitor and Remote Readback Full

 

 

 

 

Scale Calibration

 

 

 

 

Constant Voltage Zero Calibration

 

 

 

 

 

A2

Control Board

Constant Voltage Circuit

All

Resistance Programming Full Scale Calibration

 

 

(Current Source)

 

 

 

 

 

 

 

A2

Control Board

Constant Current Circuit

All

Current Monitor Zero Calibration

 

 

 

 

Constant Current Zero Calibration

 

 

 

 

Current Monitor Full Scale Calibration

 

 

 

 

Constant Current Full Scale Calibration

 

 

 

 

 

A2

Control Board

Power Limit Comparator

All

Power Limit Calibration

 

 

 

 

 

A2

Control Board

Bias Power Supplies

All

All Calibration procedures

 

 

( + & -15V Supplies)

 

 

 

 

 

 

 

A8

GPIB Board

Voltage Monitor Buffer

All

Voltage Monitor Zero Calibration

 

 

 

 

Remote Readback Zero Calibration

 

 

 

 

Constant Voltage Full Scale Calibration

 

 

 

 

Voltage Monitor and Remote Readback Full

 

 

 

 

Scale Calibration

 

 

 

 

Constant Voltage Zero Calibration

 

 

 

 

 

A8

GPIB Board

Analog Multiplexer

All

Remote Readback Zero Calibration

 

 

 

 

Constant Voltage Full Scale Calibration

 

 

 

 

Voltage Monitor and Remote Readback Full

 

 

 

 

Scale Calibration

 

 

 

 

Constant Voltage Zero Calibration

 

 

 

 

 

A8

GPIB Board

Readback DAC

All

Remote Readback Zero Calibration

 

 

 

 

Constant Voltage Full Scale Calibration

 

 

 

 

Voltage Monitor and Remote Readback Full

 

 

 

 

Scale Calibration

 

 

 

 

Constant Voltage Zero Calibration

 

 

 

 

 

13

Page 12
Image 12
Agilent Technologies 6035A, 6031A, 6032A, 6030A service manual CR4