Figure 2-8. Load Transient Recovery Waveform
g.Check that the amplitude of the transient pulse at 1 ms is no more than: 150mV/2ms (6030A)
100mV/2ms (6031A) 100mV/2ms (6032A) 200mV/5ms (6035A)
Constant Current (CC) Tests
CCSetup.
Current Programming And Readback Accuracy. This procedure verifies that the current programming and readback functions are within specifications. A GPIB controller must be used for this test. The accuracy of the current shunt resistor (Rm) must be 0.02% or better. Proceed as follows:
a.Connect test setup shown in Figure
b.Turn on ac power to the power supply.
c.Send string:
"VSET 200; ISET 0.5" (6030A) "VSET 20; ISET 0.5" (6031A) ''VSET 60; ISET 0.5" (6032A)
''VSET 500; ISET 0.10" (6035A)
d.Check that the voltage across Rm is in the range: 4.75 to 5.25mV (6030A)
248to 751∝V (6031A)
4.14 to 5.86mV (6032A)
1.5 to 1.85mV (6035A) Note the reading.
e. Enter and run the following program:
10OUTPUT 705; "IOUT?''
20 ENTER 705; A
30 DISP A
40 GOTO 10
50END
f. The value displayed by the controller should be the actual output current:
±17mA (6030A)
±102mA (6031A)
±36mA (6032A)
±50 mA (6035A)
28