On Pages
42Vdc (6030A) 5.2Vdc (6031A) 13.5Vdc (6032A) 150Vdc (6035A)
On Page 28 and 29, in paragraph Current Programming and Readback Accuracy steps c and g, change the VSET values to: VSET 170 (6030A)
VSET 50 (6032A)
VSET 425 (6035A)
On Page 29, in Load Effect (Load Regulation) step c, change the output current to: 4.7Adc (6030A)
35Adc (6032A) 16Adc (6035A) 1.38Adc (6035A)
On Page 29, in Load Effect (Load Regulation) step cd change the output voltage to: 170Vdc (6030A)
50Vdc (6032A) 425Vdc (6035A)
On Page 30, in Source Effect (Line Regulation) step d, change the output current to: 17Adc (6030A)
120Adc (6032A) 50Adc (6035A) 5Adc (6035A)
On Page 30, in Source Effect (Line Regulation) step e, change the output voltage to: 42Vdc (6030A)
5.2Vdc (6031A) 13.5Vdc (6032A) 150Vdc (6035A)
On Page 30, in PARD Ripple and Noise, make the following resistor changes: from 3.5 to 2.4 ohms (6030A)
from 0.069 to 0.043 ohms (6031A) from 0.4 to 0.27 ohms (6032A) from 40 to 30 ohms (6035A)
On Page 30, Initialization, change step 2 as follows:
2. Turn the power on, and depending on your unit's model number, send the following strings:
"EEINIT 6030" | ''EEINIT 6031'' |
"EEPROM 10,4,175.00" | ''EEPROM 10,4,17.5" |
''EEINIT 6032" | "EEINIT 6035" |
"EEPROM 10,4,51.495" | "EEPROM 10,4,437.50" |