136
Performance Test Record - Agilent 6035A POWER SUPPLY (Page 2 of 2)
MODEL Agilent 6035A Report No.______________ Date_____________________
Page
No.
Test Description Minimum Spec. Resul ts
*
Maximum
Spec.
Measurement
Uncertainty
Constant Voltage Tests
23 Voltage Programming
and Readback
Low Voltage (1.00V)
Front Panel Display
GPIB Readback
0.598
VOUT - 0.300
VOUT - 0.205
________V
________V
________V
1.400
VOUT + 0.300
VOUT + 0.205
12µV
12µV
12µV
High Voltage (500V)
Front Panel Display
GPIB Readback
498.350
VOUT - 3.500
VOUT - 2.700
________V
________V
________V
501.650
VOUT + 3.500
VOUT + 2.700
6.5mV
6.5mV
6.5mV
23 Load Effect (200V) VOUT - 0.060 _______mV VOUT + 0.060 350µV
24 Source Effect (500V) VOUT - 0.063 _______mV VOUT + 0.063 800µV
25
26
PARD (Ripple and Noise)
RMS
Peak-to-Peak
0
0
_______mV
_______mV
0.050
0.160
7.5mV
4.8mV
27 Transient Recovery
(90% to 100% Full Load)
0 _______mV 0.200/5ms 24mV
Constant Current Tests
28 Current Programming
and Readback
Low Current (0.10A)
Front Panel Display
GPIB Readback
0.015
IOUT - 0.060
IOUT - 0.050
________A
________A
________A
0.185
IOUT + 0.060
IOUT + 0.050
56µA
56µA
56µA
High Current (5.0A)
Front Panel Display
GPIB Readback
4.900
IOUT - 0.085
IOUT - 0.075
________A
________A
________A
5.100
IOUT + 0.085
IOUT + 0.075
2mA
2mA
2mA
29 Load Effect (2.0A) IOUT - 0.034 ________mA IOUT + 0.034 2µA
29 Source Effect (5.0A) IOUT - 0.018 ________mA IOUT + 0.018 2µA
30 PARD (Ripple and Noise) 0 ________mA 0.050 5mA