Performance Test Record - Agilent 6035A POWER SUPPLY (Page 2 of 2)

 

 

MODEL Agilent 6035A

Report No.______________

Date_____________________

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page

Test Description

Minimum Spec.

Results

 

Maximum

Measurement

No.

 

 

*

 

Spec.

Uncertainty

 

 

 

Constant Voltage Tests

 

 

 

 

 

23

 

Voltage Programming

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

and Readback

 

 

 

 

 

 

 

 

 

Low Voltage (1.00V)

0.598

________V

 

1.400

12∝V

 

 

Front Panel Display

VOUT - 0.300

________V

 

VOUT + 0.300

 

 

 

GPIB Readback

 

12 V

 

 

VOUT - 0.205

________V

 

VOUT + 0.205

12∝V

 

 

High Voltage (500V)

498.350

________V

 

501.650

6.5mV

 

 

Front Panel Display

VOUT - 3.500

________V

 

VOUT + 3.500

6.5mV

 

 

GPIB Readback

VOUT - 2.700

________V

 

VOUT + 2.700

6.5mV

23

 

Load Effect (200V)

VOUT - 0.060

_______mV

 

VOUT + 0.060

 

 

 

 

 

350 V

24

 

Source Effect (500V)

VOUT - 0.063

_______mV

 

VOUT + 0.063

 

 

 

 

 

800 V

 

 

PARD (Ripple and Noise)

 

 

 

 

 

 

 

25

 

RMS

0

_______mV

 

0.050

7.5mV

26

 

Peak-to-Peak

0

_______mV

 

0.160

4.8mV

27

 

Transient Recovery

0

_______mV

 

0.200/5ms

24mV

 

 

(90% to 100% Full Load)

 

 

 

 

 

 

 

 

 

 

Constant Current Tests

 

 

 

 

 

28

 

Current Programming

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

and Readback

 

 

 

 

 

 

 

 

 

Low Current (0.10A)

0.015

________A

 

0.185

56∝A

 

 

Front Panel Display

IOUT - 0.060

________A

 

IOUT + 0.060

 

 

 

GPIB Readback

 

56 A

 

 

IOUT - 0.050

________A

 

IOUT + 0.050

56∝A

 

 

High Current (5.0A)

4.900

________A

 

5.100

2mA

 

 

Front Panel Display

IOUT - 0.085

________A

 

IOUT + 0.085

2mA

 

 

GPIB Readback

IOUT - 0.075

________A

 

IOUT + 0.075

2mA

29

 

Load Effect (2.0A)

IOUT - 0.034

________mA

 

IOUT + 0.034

2

 

 

 

 

A

29

 

Source Effect (5.0A)

IOUT - 0.018

________mA

 

IOUT + 0.018

2

 

 

 

 

A

30

 

PARD (Ripple and Noise)

0

________mA

 

0.050

5mA

 

 

 

 

 

 

 

 

 

 

136

Page 131
Image 131
Agilent Technologies 6030A Performance Test Record Agilent 6035A Power Supply Page 2, Load Effect 2.0A, Source Effect 5.0A