132
Performance Test Record - Agilent 6031A POWER SUPPLY (Page 2 of 2)
MODEL Agilent 6031A Report No.______________ Date_____________________
Page
No.
Test Description Minimum Spec. Resul ts
*
Maximum
Spec.
Measurement
Uncertainty
Constant Voltage Tests
23 Voltage Programming
and Readback
Low Voltage (0.10V)
Front Panel Display
GPIB Readback
0.085
VOUT - 0.007
VOUT - 0.007
________V
________V
________V
0.115
VOUT + 0.007
VOUT + 0.007
3µV
3µV
3µV
High Voltage (20.0V)
Front Panel Display
GPIB Readback
19.978
VOUT - 0.028
VOUT - 0.023
________V
________V
________V
20.022
VOUT + 0.028
VOUT + 0.023
335µV
335µV
335µV
23 Load Effect (7.0V) VOUT - 0.0037 _______mV VOUT + 0.0037 1µV
24 Source Effect (20.0V) VOUT - 0.004 _______mV VOUT + 0.004 20µV
25
26
PARD (Ripple and Noise)
RMS
Peak-to-Peak
0
0
_______mV
_______mV
0.008
0.050
1.2mV
1.5mV
27 Transient Recovery
(90% to 100% Full Load)
0 _______mV 0.100/2ms 12mV
Constant Current Tests
28 Current Programming
and Readback
Low Current (0.50A)
Front Panel Display
GPIB Readback
0.248
IOUT - 0.303
IOUT - 0.102
________A
________A
________A
0.751
IOUT + 0.303
IOUT + 0.102
220µA
220µA
220µA
High Current (120A)
Front Panel Display
GPIB Readback
119.450
IOUT - 1.140
IOUT - 0.580
________A
________A
________A
120.550
IOUT + 1.140
IOUT + 0.580
50mA
50mA
50mA
29 Load Effect (50.0A) IOUT - 0.020 ________mA IOUT + 0.020 25µA
29 Source Effect (120A) IOUT - 0.037 ________mA IOUT + 0.037 136mA
30 PARD (Ripple and Noise) 0 ________mA 0.120 18mA