Figure 2-9. CC PARD Test Setup

Initialization Procedure

Follow the procedure if either the GPIB assembly has been replaced, or the EEPROM (U70) has been replaced:

1.Install the GPIB assembly in the unit.

2.Turn the power on and depending on your unit's model number, send string: "EEINIT 6030"

"EEINIT 6031'' "EEINlT 6032" or

''EEINIT 6035"

3.Turn the power off, wait 5 seconds, then turn the power back on.

4.If the GPIB assembly has been replaced, calibrate the unit.

31