Table 3-8. Secondary Processor Signature Table

DS(0)

P36U

U4-1

U7-15

U9-15

U11-15

 

DS(1)

2280

U4-2

U7-14

U9-14

U11-14

 

DS(2)

4277

U4-3

U7-13

U9-13

U11-13

 

DS(3)

720F

U4-4

U7-12

U9-12

U11-12

 

DS(4)

6A31

U4-5

U7-11

U9-11

U11-11

 

DS(5)

662U

U4-6

U7-10

U9-10

U11-10

 

DS(6)

6020

U4-7

U7-9

U9-9

U11-9

 

DS(7)

6327

U4-8

U7-8

U9-8

U11-8

 

DS(8)

1377

U4-39

U7-7

U9-7

U11-7

U2-16

DS(9)

FF99

U4-38

U7-6

U9-6

U11-6

U2-17

DS(10)

236P

U4-37

U7-5

U9-5

U11-5

U2-18

DS(11)

H495

U4-36

U7-4

U9-4

U11-4

U2-19

WR*

9FU7

U4-12

U7-17

 

 

 

WR*

9FF7

U4-13

 

U9-17

 

 

WR*

9FHU

U4-14

 

 

U11-17

 

 

F817

U4-26

 

 

 

U20-1

 

36U7

U4-27

 

 

 

U20-16

 

0562

U4-28

 

 

 

U20-15

ISTX

9F97

U4-11

 

 

 

U2-4, 14

 

9FH6

U4-17

 

 

 

U2-2

 

9FH5

U4-16

 

 

 

U2-11

ALE

0000

U4-30

 

 

 

U2-1

 

AU68

U4-35

 

 

 

 

+ 5V

9FA8

U4-33,34

 

 

 

 

Power Section Troubleshooting

Table 3-9 describes the signals at each of the control board test points. The test connector provided in service kit P/N

5060-2865 allows easy connection to each test point. The measurements given here include bias and reference voltages as well as power supply status signals. It provides conditions for these measurements and gives the components which are the sources of the signals. Tables 3-10 and 3-11 describe possible symptoms in the power section. Both give lists of circuit blocks or components which can cause the symptoms shown. The appropriate assembly is also given.

If the supply exhibits a symptom given in Table 3-10 or 3-11, go to the block which pertains to that symptom. If the exact symptom seen is not in the tables, start with the symptom that seems to be closest to the one observed. The blocks are given in the Power Section Blocks section starting in Paragraph 3-84. Troubleshooting information for each block will include a brief description of the circuit. The columns provided are as follows:

NODE:

This column lists the nodes where the measurements should be taken. In some cases this will be

 

stated as NODE( + ) and NODE(-) where the first is the test node and the second is the reference.

SETUP:

If a certain setup is required for the measurement, it will be given in this column.

MEASUREMENT:

This column indicates what the expected measurement is for the given node.

SOURCE:

If applicable, the components which generate the signal will be provided in this column.

51

Page 49
Image 49
Agilent Technologies 6031A, 6035A, 6032A, 6030A Power Section Troubleshooting, Secondary Processor Signature Table