130
Performance Test Record - Agilent 6030A POWER SUPPLY (Page 2 of 2)
MODEL Agilent 6030A Report No.______________ Date_____________________
Page
No.
Test Description Minimum Spec. Resul ts
*
Maximum
Spec.
Measurement
Uncertainty
Constant Voltage Tests
23 Voltage Programming
and Readback
Low Voltage (0.500V)
Front Panel Display
GPIB Readback
0.354
VOUT - 0.065
VOUT - 0.080
________V
________V
________V
0.645
VOUT + 0.065
VOUT + 0.080
7µV
7µV
7µV
High Voltage (200V)
Front Panel Display
GPIB Readback
199.785
VOUT - 0.72
VOUT - 0.24
________V
________V
________V
200.215
VOUT + 0.72
VOUT + 0.24
2.7mV
2.7mV
2.7mV
23 Load Effect (60.0V) VOUT - 0.011 _______mV VOUT + 0.011 40µV
24 Source Effect (200V) VOUT - 0.011 _______mV VOUT + 0.011 350µV
25
26
PARD (Ripple and Noise)
RMS
Peak-to-Peak
0
0
_______mV
_______mV
0.022
0.050
3.3mV
1.5mV
27 Transient Recovery
(90% to 100% Full Load)
0 _______mV 0.150/2ms 18mV
Constant Current Tests
28 Current Programming
and Readback
Low Current (0.50A)
Front Panel Display
GPIB Readback
0.475
IOUT - 0.017
IOUT - 0.017
________A
________A
________A
0.525
IOUT + 0.017
IOUT + 0.017
220µA
220µA
220µA
High Current (17.0A)
Front Panel Display
GPIB Readback
16.941
IOUT - 0.081
IOUT - 0.076
________A
________A
________A
17.059
IOUT + 0.081
IOUT + 0.076
7mA
7mA
7mA
29 Load Effect (5.0A) IOUT - 0.010 ________mA IOUT + 0.010 2µA
29 Source Effect (17.0A) IOUT - 0.006 ________mA IOUT + 0.006 15µA
30 PARD (Ripple and Noise) 0 ________mA 0.015 2mA