Chapter 3 Manual Changes:

On Page 45, in Readback Multiplexer (U20), change Node U20-9, Measurement from + 5V to:

+4.25V (6030A/31A/35A)

+4.167V (6032A)

On Page 53 in Main Troubleshooting and on Page 54 in Troubleshooting No-Out Failures step d, change 320Vdc to 250Vdc.

Chapter 4 Manual Changes:

On Page 74 In Power Mesh and Input Circuits, change 320Vdc to 250Vdc.

Chapter 5 and 6 Manual Changes

On Page 86, change the A2 board from: 06030-60022 to 06030-60026 (6030A) 06011-60022 to 06031-60026 (6031A) 06032-60022 to 06032-60026 (6032A) 06030-60028 to 06035-60020 (6035A)

On Page 91 and on the schematic diagram Figure 6-10, change R81 to: 1.82K, 1% 1/8W Agilent P/N 0757-0429 (6030A, 6035A only) 1.78K, 1% 1/8W Agilent P/N 0757-0278 (6032A only)

On Page 92 and on the schematic diagram Figure 6-10, change R92 to: 10K, 1% 1/8W Agilent P/N 0757-0442 (6030A, 6035A only) 34.8K, 1% 1/8W Agilent P/N 0757-0123 (6031A only)

11.5K, 1% 1/8W Agilent P/N 0698-5383 (6032A only)

On Page 94, add option 100 label (A2 board) 9320-5540.

On Page 102, change front panel from: 06030-00012 to 06030-00011 (6030A) 06031-00013 to 06031-00012 (6031A) 06032-00023 to 06032-00022 (6032A) 06035-00001 to 06035-00003 (6035A)

On Page 103, add line voltage label (rear panel) Agilent P/N 06032-81004.

123