Performance Tests

Agilent E-Series Power Sensor Functional Test

Agilent E-Series Power Sensor Functional Test

Description

This test verifies that the meter/sensor combination can make RF measurements.

The Agilent E-series power sensors operate over a 90 dB dynamic range (-70 to +20 dBm). However, since there is an amplifier in these power sensors, the voltages presented to the power sensor connector are always within the range of voltages available from the Agilent 11683A range calibrator.

Equipment

Power Sensor Cable

Agilent 11730A

Reference Attenuator

Agilent 11708A

Agilent E-series power sensor

 

Procedure

To complete the Functional Test the following procedure should be performed.

Note

The procedure details the key presses required on the

 

Agilent E4418B. For the Agilent E4419B the equivalent key

 

presses should be performed on both channels.

 

 

 

1. Connect the equipment as shown in Figure 2-5.

Power Meter

Agilent E-Series

 

Power Sensor

 

POWER

 

REF

 

CHANNEL A

 

Agilent 11730A

Figure 2-5: Sensor Functional Test Setup

2-20

Agilent E4418B/E4419B Service Guide

Page 54
Image 54
Agilent Technologies e4418b, e4419b manual Agilent E-Series Power Sensor Functional Test, Sensor Functional Test Setup