Performance Tests
Instrument Accuracy Test
Test Setup
Power Meter
Range
Calibrator
CHANNEL A | POWER |
| METER |
| Agilent 11730A |
Figure 2-2: Instrument Accuracy Test Setup
Procedure
The following procedure should be performed for the Instrument Accuracy Test.
Note | The procedure details the key presses required on the | ||||||||||
| Agilent E4418B. For the Agilent E4419B the equivalent key | ||||||||||
| presses should be performed on both channels. | ||||||||||
|
|
|
|
|
|
|
|
|
|
| |
| 1. | Connect the equipment as shown in Figure | |||||||||
| 2. | Switch the power meter on. | |||||||||
| 3. | Press |
|
| then |
| . | ||||
| Preset | ||||||||||
|
| Confirm | |||||||||
| Local | ||||||||||
|
|
| |||||||||
| 4. | Press |
| dBm/W |
| , |
| for a reading in watts. | |||
|
|
| W | ||||||||
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
5. Set the range calibrator as follows:
RANGE | 3 ∝W |
POLARITY | NORMAL |
FUNCTION | STANDBY |
LINE | ON |
Agilent E4418B/E4419B Service Guide |