Figure 2-2: Instrument Accuracy Test Setup

Performance Tests

Instrument Accuracy Test

Test Setup

Power Meter

Range

Calibrator

CHANNEL A

POWER

 

METER

 

Agilent 11730A

Figure 2-2: Instrument Accuracy Test Setup

Procedure

The following procedure should be performed for the Instrument Accuracy Test.

Note

The procedure details the key presses required on the

 

Agilent E4418B. For the Agilent E4419B the equivalent key

 

presses should be performed on both channels.

 

 

 

 

 

 

 

 

 

 

 

 

1.

Connect the equipment as shown in Figure 2-2.

 

2.

Switch the power meter on.

 

3.

Press

 

 

then

 

.

 

Preset

 

 

Confirm

 

Local

 

 

 

 

4.

Press

 

dBm/W

 

,

 

for a reading in watts.

 

 

 

W

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5. Set the range calibrator as follows:

RANGE

3 ∝W

POLARITY

NORMAL

FUNCTION

STANDBY

LINE

ON

2-8

Agilent E4418B/E4419B Service Guide

Page 42