Adjustments
Power Reference Oscillator Frequency Adjustment
3. Turn the power reference on by pressing
■ ZerCalo , Power Ref Off On on the Agilent E4418B.
■ZerCalo , Cal, More , Power Ref Off On on the Agilent E4419B.
4.Observe the reading on the frequency counter. If it is
50.0 ±0.5 MHz, no adjustment of the power reference oscillator frequency is necessary. If it is not within these limits, adjust the power reference oscillator frequency as described in step 5 and step 6.
5.Remove the power meter cover.
6.Adjust A2L9 to obtain a 50.0 ±0.5 MHz indication on the frequency counter. Refer to Figure
A2R90
BT1
A2 Assembly | A2L9 |
A2J13 J12
Front Panel
Figure 3-2: A2L9 Adjustment Location
Agilent E4418B/E4419B Service Guide |