Troubleshooting
Introduction
A
Select
Preset/local
Confirm
Meas/Setup
Limits
Limits On
TTL Output
TTL Output On, Fail O/P Low
DVM reading
Yes
Select
Fail O/P High
No
Power down meter.
Disconnect ac power and battery if fitted. Remove A2 Processor assy, as directed on page
DVM reading
Yes
No Fault found with Upper Window TTL Over Limit detection circuit - check user application/implementation
No
A2 Processor Assy
Defective
Yes |
| DVM reading |
| No | |
| Ω | +/- 10% |
| ||
| 450 |
|
|
A4 Comms Assy
Defective
End
Agilent E4418B/E4419B Service Guide |