Performance Tests

Zero Test

Zero Test

Specification

Electrical Characteristics

Performance Limits

 

 

 

 

Accuracy: Zero set (Digital settability of zero)

±0.0764 ∝W1

1.This performance limit is determined by the zero set specification of the power

sensor used in the measurement plus the measurement noise. The range calibrator has a zero set specification of ±0.05 ∝W. The calibrator measurement

noise specification is 110 nW at 16 averages. At 512 averages a noise multiplier of 0.24 is required, giving a measurement noise specification of 0.0264 ∝W (0.24x512).

Description

After the power meter is initially calibrated and zeroed, the change in the digital readout is monitored. This test also takes drift and noise into account, since drift, noise and zero readings cannot be separated.

Equipment

Range Calibrator

Agilent 11683A

Power Sensor Cable

Agilent 11730A

Test Setup

Power Meter

Range

Calibrator

CHANNEL A

POWER

 

METER

 

Agilent 11730A

Figure 2-1: Zero Test Setup

Agilent E4418B/E4419B Service Guide

2-5

Page 39
Image 39