Agilent Technologies e4418b, e4419b manual Zero Drift, Measurement Noise

Models: e4419b e4418b

1 144
Download 144 pages 36.25 Kb
Page 24
Image 24
Zero Drift1

Specifications

Power Meter Supplemental Characteristics

Table 1-3: Power Sensor Specifications±

Power Sensor

Zero Drift1

Measurement Noise2

 

 

 

Agilent 8481A

<±10 nW

<110 nW

 

 

 

Agilent 8481B

<±10 W

<110 W

 

 

 

Agilent 8481D

<±4 pW

<45 pW

 

 

 

Agilent 8481H

<±1 W

<10 W

 

 

 

Agilent 8482A

<±10 nW

<110 nW

 

 

 

Agilent 8482B

<±10 W

<110 W

 

 

 

Agilent 8482H

<±1 W

<10 W

 

 

 

Agilent 8483A

<±10 nW

<110 nW

 

 

 

Agilent 8485A

<±10 nW

<110 nW

 

 

 

Agilent 8485D

<±4 pW

<45 pW

 

 

 

Agilent R8486A

<±10 nW

<110 nW

 

 

 

Agilent R8486D

<±6 pW

<65 pW

 

 

 

Agilent Q8486A

<±10 nW

<110 nW

 

 

 

Agilent Q8486D

<±6 pW

<65 pW

 

 

 

Agilent W8486A

<±40 nW

<450 nW

 

 

 

Agilent 8487A

<±10 nW

<110 nW

 

 

 

Agilent 8487D

<±4 pW

<45 pW

 

 

 

Agilent E4412A

15 pW

<70 pW

 

 

 

Agilent E4413A

15 pW

<70 pW

 

 

 

1.Within 1 hour after zero set, at a constant temperature, after a 24 hour warm-up of the power meter.

2.The number of averages at 16 (for normal mode) and 32 (for x2 mode), at a constant temperature, measured over a 1 minute interval and 2 standard deviations. For Agilent E-series power sensors the measurement noise is measured within the low range. Refer to the relevant power sensor manual for further information.

1-8

Agilent E4418B/E4419B Service Guide

Page 24
Image 24
Agilent Technologies e4418b, e4419b manual Zero Drift, Measurement Noise