| Specifications |
|
| Power Meter Specifications |
1 mW Power Reference1 |
|
|
• Power Output: | 1.00 mW (0.0 dBm). | |
• Accuracy: | (for two years) | |
| ±0.9% | (0 to 55 ºC). |
| ±0.6% | (25±10 ºC). |
| ±0.5% | (23±3 ºC). |
• Frequency: | 50 MHz nominal. | |
• SWR2: | 1.06 maximum (1.08 maximum for | |
| option 003). | |
• Connector Type: | Type N (f), 50 ohms. |
1.National metrology institutes of member states of the Metre Convention, such as the National Institute of Standards and Technology in the USA, are signatories to the Comité International des Poids et Mesures Mutual Recognition Arrangement. Further information is available from the Bureau International des Poids et Mesures, at http://www.bipm.fr/
2.This SWR specification is only warranted for the E4418B and the E4419B Power Meters with serial prefix GB4331xxxx and above. Prior to this prefix the values shown are supplemental characteristics.
Agilent E4418B/E4419B Service Guide |