Performance Tests

 

Zero Test

 

Procedure

 

The following procedure should be performed for the Zero Test.

 

 

Note

The procedure details the key presses required on the

 

Agilent E4418B. For the Agilent E4419B the equivalent key

 

presses should be performed on both channels.

 

 

1.Connect the equipment as shown in Figure 2-1.

2.Switch the power meter on.

3.Press PreseLocal t then Confirm .

4.Press dBm/W , W for a reading in watts.

5.Set the range calibrator as follows:

RANGE

3 ∝W

POLARITY

NORMAL

FUNCTION

STANDBY

LINE

ON

Note

When switching the range calibrator to STANDBY, allow enough

 

 

time for the range calibrator to settle to its zero value before

 

 

attempting to zero the power meter. This settling would appear on

 

the power meter display as downward drift. When the drift has

 

 

reached minimum, (typically less than 60 seconds), the range

 

 

calibrator is settled.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6. On the power meter press

System

,

 

 

,

 

 

,

 

 

Input Settings

 

 

 

 

More

 

 

Inputs

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Filter , Filter On , Mode Man , Length , and set the filter length to 512 in the pop up window then press Enter .

7.On the power meter press ZerCalo , Zero. Wait approximately 10 seconds for the wait symbol to disappear.

8.Wait 30 seconds then verify that the power meter’s reading is within ±0.05 ∝W. Record the reading.

Table 2-2: Zero Test Results

 

 

 

Channel B

 

 

Min

Channel A

Actual Result

Max

 

Actual Result

(Agilent E4419B

 

 

 

 

 

 

Only)

 

 

 

 

 

 

 

 

 

 

 

 

-76.40 nW

________________

________________

+76.40 nW

 

 

 

 

 

 

 

 

 

 

2-6

Agilent E4418B/E4419B Service Guide

Page 40
Image 40
Agilent Technologies e4418b, e4419b manual Procedure, Range Polarity Normal Function Standby Line