| Performance Tests |
| Zero Test |
| Procedure |
| The following procedure should be performed for the Zero Test. |
|
|
Note | The procedure details the key presses required on the |
| Agilent E4418B. For the Agilent E4419B the equivalent key |
| presses should be performed on both channels. |
|
|
1.Connect the equipment as shown in Figure
2.Switch the power meter on.
3.Press PreseLocal t then Confirm .
4.Press dBm/W , W for a reading in watts.
5.Set the range calibrator as follows:
RANGE | 3 ∝W |
POLARITY | NORMAL |
FUNCTION | STANDBY |
LINE | ON |
Note | When switching the range calibrator to STANDBY, allow enough |
| |||||||
| time for the range calibrator to settle to its zero value before |
| |||||||
| attempting to zero the power meter. This settling would appear on | ||||||||
| the power meter display as downward drift. When the drift has |
| |||||||
| reached minimum, (typically less than 60 seconds), the range |
| |||||||
| calibrator is settled. |
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 6. On the power meter press | System | , |
|
| , |
|
| , |
|
| Input Settings |
|
| |||||
|
| More |
| ||||||
| Inputs |
|
| ||||||
|
|
| |||||||
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
Filter , Filter On , Mode Man , Length , and set the filter length to 512 in the pop up window then press Enter .
7.On the power meter press ZerCalo , Zero. Wait approximately 10 seconds for the wait symbol to disappear.
8.Wait 30 seconds then verify that the power meter’s reading is within ±0.05 ∝W. Record the reading.
Table
|
|
| Channel B |
|
| Min | Channel A | Actual Result | Max |
| Actual Result | (Agilent E4419B | ||
|
|
| ||
|
|
| Only) |
|
|
|
|
|
|
|
|
|
|
|
| ________________ | ________________ | +76.40 nW | |
|
|
|
|
|
|
|
|
|
|
Agilent E4418B/E4419B Service Guide |