Performance Tests

Performance Tests

Introduction

 

Introduction

 

The procedures in this chapter test the power meter’s electrical

 

performance using the specifications in Chapter 1 as the performance

 

standards. All tests can be performed without access to the interior of the

 

power meter. A simpler operational test is included in “Self Test” of the

 

User’s Guide.

 

 

Note

For valid performance tests, the following conditions must be met:

 

a. The power meter and test equipment must have a 30 minute

 

warm-up for all specifications.

 

b. The line voltage for the power meter must be 85 to 264 Vac and

 

the line frequency must be 50 to 440 Hz.

 

c. The ambient temperature must be 00 to 550C.