Performance Tests
Introduction
| Introduction |
| The procedures in this chapter test the power meter’s electrical |
| performance using the specifications in Chapter 1 as the performance |
| standards. All tests can be performed without access to the interior of the |
| power meter. A simpler operational test is included in “Self Test” of the |
| User’s Guide. |
|
|
Note | For valid performance tests, the following conditions must be met: |
| a. The power meter and test equipment must have a 30 minute |
| |
| b. The line voltage for the power meter must be 85 to 264 Vac and |
| the line frequency must be 50 to 440 Hz. |
| c. The ambient temperature must be 00 to 550C. |
Agilent E4418B/E4419B Service Guide |