Troubleshooting

Introduction

A

Select

Preset/local

Confirm

Lower Window - using

Meas/Setup

Limits

Limits On

TTL Output

TTL Output On, Fail O/P Low

 

 

 

Yes

 

DVM reading

 

 

No

 

 

 

 

 

 

 

5.0V+/-0.5V?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Power down meter.

Select

 

 

 

 

 

 

Disconnect ac power and battery if fitted.

Fail O/P High

 

 

 

Remove A2 Processor assy, as directed on page

 

 

 

 

 

 

 

5-9, to access the Comms Assembly connector.

 

 

 

 

 

 

 

Use DVM to check resistance between J4 pin 42

 

 

 

 

 

 

 

and Rmt I/O connector pin 4.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DVM reading

Yes 0.0V+/-0.5V?

No Fault found with Upper Window TTL Over Limit detection circuit - check user application/implementation

No

A2 Processor Assy

Defective

Yes

 

DVM reading

 

No

 

450+/- 10%

 

 

 

A4 Comms Assy

Defective

End

6-6

Agilent E4418B/E4419B Service Guide

Page 114
Image 114