Agilent Technologies e4419b manual Post-Repair Adjustments, Equipment Required, Assembly Replaced

Models: e4419b e4418b

1 144
Download 144 pages 36.25 Kb
Page 61
Image 61
Post-Repair Adjustments

Adjustments

Introduction

Equipment Required

The adjustment procedures include a list of recommended test equipment. The test equipment is also identified on the test setup diagrams.

Post-Repair Adjustments

Table 3-1 lists the adjustments related to repairs or replacement of any of the assemblies.

Table 3-1: Post Repair Adjustments, Tests, and Checks

 

Assembly Replaced

Related Adjustments, Performance

 

Tests or Self Tests

 

 

 

 

 

 

 

 

A1

Power Supply

Self Test.

 

 

 

A2

Processor Assembly

Frequency and Level Adjustments,

 

 

Power Reference Level Test, Display

 

 

Brightness and Contrast Adjustment.

 

 

 

A3

Front Panel Assembly

Instrument Accuracy Test, Display

 

 

Brightness and Contrast Adjustment.

 

 

 

A4

Comms Assembly

All automated Tests and Serial Interface

 

 

Self Test.

 

 

 

A5

Daughter Assembly

Instrument Accuracy Test.

 

 

A6 Measurement Assembly

Zero Test, Instrument Accuracy Test.

 

 

W1 or W2 Power Sensor

Instrument Accuracy Test, Zero Test.

Cable

 

 

 

BT1 Lithium Battery

Battery Self Test.

 

 

Power Reference Cable Kits

Power Reference Level Test.

 

 

 

Agilent E4418B/E4419B Service Guide