Agilent E-Series Power Sensor Interface Test

Performance Tests

Agilent E-Series Power Sensor Interface Test

Agilent E-Series Power Sensor Interface Test

The Agilent E-series power sensors have their sensor calibration tables stored in EEPROM which enables the frequency and calibration factor data to be downloaded automatically by the power meter. The frequency and calibration factor data have checksums which are compared with the data downloaded by the power meter. For the Agilent E4419B dual channel power meter, this test should be performed on both channels.

Procedure

1.When there is no power sensor connected to the Channel A input the message “No Sensor” (“No Sensor ChA” on the

Agilent E4419B) appears on the measurement window.

2.Connect the Agilent E-series power sensor to the power meter’s Channel A input. The message “Reading Sensor ChA” appears. When all the data is downloaded this message disappears and a reading is displayed on the measurement window. If the download fails, a warning message and error flag are displayed on the front panel. Also, SCPI error message “-310, System Error” is generated.

3.For the Agilent E4419B repeat these steps for Channel B.

Agilent E4418B/E4419B Service Guide