Theory of Operation

A5 Daughter Assembly

A5 Daughter Assembly

The A5 daughter assembly is loaded vertically into the A2 processor assembly. The Agilent E4418B has five connectors on the A5 daughter assembly:

two 6-way connectors route the signal lines between the rechargeable battery assembly and the A1 power supply/charger assembly, and provide an interface for the signal lines that are monitored by the A2 processor assembly.

a 48-way connector provides the interface for the A2 processor assembly signals.

a 36-way slotted post connector provides the interface for the A6 measurement assembly.

a 2-way connector for production test only.

The Agilent E4419B has an additional 36-way slotted post connector.