Performance Tests
Introduction
Equipment Required
Table
Table 2-1: Required Equipment
Equipment | Model Number | Critical Specification | Usage1 | |
Required | ||||
|
|
| ||
|
|
|
| |
Digital Multimeter | Agilent 3458A | Range 0 to 20 V | P and A | |
|
| Resolution 0.01 V |
| |
|
|
|
| |
DC Calibrator | Datron 4000A |
| P | |
| or |
|
| |
| Fluke 5720A |
|
| |
|
|
|
| |
Range Calibrator | Agilent 11683A | Option H01 is essential for | P | |
| or | accurate verification. (See |
| |
| Agilent 11683A | analysis in Appendix A). The |
| |
| Opt H01 | Standard 11683A can be used |
| |
|
| for indication only. |
| |
|
|
|
| |
Frequency Counter | Agilent 53132A | Range 10 Hz to 50 MHz | P and A | |
|
| Resolution 1 Hz |
| |
|
|
|
| |
Power Meter | Agilent 432A | Range 1 mW | P and A | |
|
| Transfer Accuracy 0.2% |
| |
|
|
|
| |
Thermistor Mount | Agilent 478A Opt H75/H76 | SWR<1.05 at 50 MHz | P and A | |
| or | Accuracy ±0.2% at 50 MHz. |
| |
| Agilent 8478B Opt H01 |
|
| |
|
|
|
| |
Power Sensor | Agilent |
| P | |
| sensor |
|
| |
|
|
|
| |
30 dB Attenuator | Agilent 11708A |
| P | |
| or |
|
| |
| Agilent 8491A Option 030 |
|
| |
|
|
|
| |
Power Sensor | Agilent 11730A |
| P and A | |
Cable | or |
|
| |
| Agilent E9288A |
|
| |
|
|
|
|
1. P = Performance Tests, A = Adjustments
Agilent E4418B/E4419B Service Guide |