Equipment Required

Performance Tests

Introduction

Equipment Required

Table 2-1 lists all the equipment required for the adjustments and performance tests. If substitutions must be made, the equipment used must meet the critical specifications.

Table 2-1: Required Equipment

Equipment

Model Number

Critical Specification

Usage1

Required

 

 

 

 

 

 

 

Digital Multimeter

Agilent 3458A

Range 0 to 20 V

P and A

 

 

Resolution 0.01 V

 

 

 

 

 

DC Calibrator

Datron 4000A

 

P

 

or

 

 

 

Fluke 5720A

 

 

 

 

 

 

Range Calibrator

Agilent 11683A

Option H01 is essential for

P

 

or

accurate verification. (See

 

 

Agilent 11683A

analysis in Appendix A). The

 

 

Opt H01

Standard 11683A can be used

 

 

 

for indication only.

 

 

 

 

 

Frequency Counter

Agilent 53132A

Range 10 Hz to 50 MHz

P and A

 

 

Resolution 1 Hz

 

 

 

 

 

Power Meter

Agilent 432A

Range 1 mW

P and A

 

 

Transfer Accuracy 0.2%

 

 

 

 

 

Thermistor Mount

Agilent 478A Opt H75/H76

SWR<1.05 at 50 MHz

P and A

 

or

Accuracy ±0.2% at 50 MHz.

 

 

Agilent 8478B Opt H01

 

 

 

 

 

 

Power Sensor

Agilent E-series power

 

P

 

sensor

 

 

 

 

 

 

30 dB Attenuator

Agilent 11708A

 

P

 

or

 

 

 

Agilent 8491A Option 030

 

 

 

 

 

 

Power Sensor

Agilent 11730A

 

P and A

Cable

or

 

 

 

Agilent E9288A

 

 

 

 

 

 

1. P = Performance Tests, A = Adjustments

Agilent E4418B/E4419B Service Guide