Manual background

Troubleshooting

Introduction

B

Remove A2 Processor Assy (see chapter 5) to gain access to the Communications Assembly connector.

Measure resistance between Rmt I/O connector pin 6 and connector J4 pin 43.

Meter reading

No

470Ω +/− 10%

Yes

Measure resistance between Rmt

I/O connector pin 5 and connector

J4 pin 44.

Meter reading

 

No

470Ω +/− 10%

 

 

 

Yes

A2 Processor assy defective

A4 Comms assy defective

End