Specifications

Power Meter Specifications

Accuracy

Instrumentation

Absolute: ±0.02 dB (Logarithmic) or ±0.5% (Linear). (Refer to the power sensor linearity specification in your power sensor manual to assess overall system accuracy.)

Relative: ±0.04 dB (Logarithmic) or ±1.0% (Linear). (Refer to the power sensor linearity specification in your power sensor manual to assess overall system accuracy.)

Zero Set (digital settability of zero): Power sensor dependent (refer to Table 1-1). For Agilent E-series power sensors, this specification applies when zeroing is performed with the sensor input disconnected from the

POWER REF.

Table 1-1: Zero Set Specifications

Power Sensor

Zero Set

 

 

 

 

Agilent 8481A

±50 nW

 

 

Agilent 8481B

±50 ∝W

 

 

Agilent 8481D

±20 pW

 

 

Agilent 8481H

±5 ∝W

 

 

Agilent 8482A

±50 nW

 

 

Agilent 8482B

±50 ∝W

 

 

Agilent 8482H

±5 ∝W

 

 

Agilent 8483A

±50 nW

 

 

Agilent 8485A

±50 nW

 

 

Agilent 8485D

±20 pW

 

 

Agilent R8486A

±50 nW

 

 

Agilent R8486D

±30 pW

 

 

Agilent Q8486A

±50 nW

 

 

Agilent Q8486D

±30 pW

 

 

Agilent V8486A

±200 nW

 

 

Agilent W8486A

±200 nW

 

 

Agilent 8487A

±50 nW

 

 

Agilent 8487D

±20 pW

 

 

Agilent E4412A

±50 pW

 

 

Agilent E4413A

±50 pW

 

 

1-4

Agilent E4418B/E4419B Service Guide