Specifications
Power Meter Specifications
Accuracy
Instrumentation
Absolute: ±0.02 dB (Logarithmic) or ±0.5% (Linear). (Refer to the power sensor linearity specification in your power sensor manual to assess overall system accuracy.)
Relative: ±0.04 dB (Logarithmic) or ±1.0% (Linear). (Refer to the power sensor linearity specification in your power sensor manual to assess overall system accuracy.)
Zero Set (digital settability of zero): Power sensor dependent (refer to Table
POWER REF.
Table 1-1: Zero Set Specifications
Power Sensor | Zero Set |
|
|
|
|
Agilent 8481A | ±50 nW |
|
|
Agilent 8481B | ±50 ∝W |
|
|
Agilent 8481D | ±20 pW |
|
|
Agilent 8481H | ±5 ∝W |
|
|
Agilent 8482A | ±50 nW |
|
|
Agilent 8482B | ±50 ∝W |
|
|
Agilent 8482H | ±5 ∝W |
|
|
Agilent 8483A | ±50 nW |
|
|
Agilent 8485A | ±50 nW |
|
|
Agilent 8485D | ±20 pW |
|
|
Agilent R8486A | ±50 nW |
|
|
Agilent R8486D | ±30 pW |
|
|
Agilent Q8486A | ±50 nW |
|
|
Agilent Q8486D | ±30 pW |
|
|
Agilent V8486A | ±200 nW |
|
|
Agilent W8486A | ±200 nW |
|
|
Agilent 8487A | ±50 nW |
|
|
Agilent 8487D | ±20 pW |
|
|
Agilent E4412A | ±50 pW |
|
|
Agilent E4413A | ±50 pW |
|
|
Agilent E4418B/E4419B Service Guide |